Skip to the content

FDA Medical Device Recalls (openFDA)

Merit Medical Systems, Inc.: VACLOCK Syringe, 20ml, REF: VAC120E, Sterile EO, Single Use, Merit Medical Systems, Inc., South Jordan, Utah 84095.

Recall number
Z-0504-2011
Recalling firm
Merit Medical Systems, Inc.
Product
VACLOCK Syringe, 20ml, REF: VAC120E, Sterile EO, Single Use, Merit Medical Systems, Inc., South Jordan, Utah 84095.
Status
Terminated
Initiated
2010-10-22
Posted
2010-12-02
Terminated
2011-01-18
Reason
Packaging for syringes may have holes which would compromise the sterility of the package contents.
Root cause
Packaging process control
Action
Merit Medical notified product managers of the product recall on October 22, 2010 following a voice mail script. Frequently Asked Questions, Sales Rep Letter, and Recall Notification Forms were then sent.. Sales reps were instructed to contact all of their customers and have them identify and quarantine any unused inventory and either return it to Merit or discard. A response form was to completed by the sales rep and the firm's representative, indicating amount of product quarantined, returned , or destroyed. Also, on 10/22/2010, authorized distributors were contacted by phone, mail, and / or fax. They were also provided a Distributor Letter, FAQs, and Recall Notification Forms instructing them to identify and quarantine any unused inventory and return it to Merit or discard. They were instructed to continue the recall to customers to whom they had shipped the affected Merit product. Additional information is available from Merit at 801-316-4822.
Quantity
789 units
Distribution
Worldwide Distribution: USA including the states of AL, CA, GA, OH, PA, TX, UT, and WA, and to the countires of Japan and UK.
Lot, serial or model codes
Lot Number: H162820
510(k) numbers
["K994253"]
PMA numbers
not on file
Product code
FMF
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register