Stille-Sonesta, Inc: Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden.
- Recall number
- Z-0504-2007
- Recalling firm
- Stille-Sonesta, Inc
- Product
- Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden.
- Status
- Terminated
- Initiated
- 2006-11-09
- Posted
- 2007-02-15
- Terminated
- 2007-06-20
- Reason
- Bolts may back out of their joints causing the back section of the table to tilt.
- Root cause
- Other
- Action
- Notifications letters sent to consignees on 11/20/06. Users are instructed to check the bolts located under the table. The Notice includes instructions with pictures for checking the bolts and tightening them if necessary. Stille AB has developed a kit for replacing the bolt that is involved and will send a service engineer out to the site to do the replacement.
- Quantity
- 66 units.
- Distribution
- Nationwide.
- Lot, serial or model codes
- Serial Numbers: 12-70000 - 12-70017, 00010, 00012, 00013 - 00022, 00024 - 00027, 00030 - 00034, 00036 - 00040, 00043 - 00044, 00045 - 00049, 00053 - 00054, 00057 - 00065, 00067 - 00069, 00071. Outside U.S. Serial Numbers: 12-70018 - 12-70019, 23-70000 - 23-70009, 23-70008, 23-700010, 00011, 00023, 00028 - 00029, 00050 - 00052, 00055 - 00056, 00066, 00072, 00073, 00074.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- IXR
- Firm FEI
- 1000516389
- Firm city
- Dallas
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28