Skip to the content

FDA Medical Device Recalls (openFDA)

NxStage Medical, Inc.: NxStage PureFlow B Solution-RFP-403, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.

Recall number
Z-0503-2019
Recalling firm
NxStage Medical, Inc.
Product
NxStage PureFlow B Solution-RFP-403, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
Status
Terminated
Initiated
2018-10-17
Posted
not on file
Terminated
2020-08-26
Reason
PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin
Root cause
Packaging process control
Action
NxStage initiated a recall on October 17, 2018 via " URGENT MEDICAL DEVICE REMOVAL" Recall Notice to its customers. The notice provided reason for recall, health risk, action to be taken. The customers were instructed to do the following: 1 Check all boxes in your Pureflow B solution inventory, separate them from your other inventory and do not use them. 2. Follow the instructions on the attached reply form or click on the following link to respond electronically, https://www.nxstage.com/rfprecalloct2018, to acknowledge your receipt of this notice. 3. Contact NxStage Customer Service to arrange for return of all affected product and for replacement product to be sent, If you have any questions or comments, please feel free to contact NxStage Customer Service 1-866- NXST AGE (1-866-697-8243).
Quantity
2189
Distribution
US Nationwide Distribution.
Lot, serial or model codes
Lot Numbers: Q1802398, Q1802428
510(k) numbers
["K053286"]
PMA numbers
not on file
Product code
KPO
Firm FEI
3003464075
Firm city
Lawrence
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register