Skip to the content

FDA Medical Device Recalls (openFDA)

TECHNO-PATH MANUFACTURING LTD.: Multichem ID-B. Model Number: SR103B. Intended for use as an unassayed positive qualitative quality control serum to monitor the precision of laboratory testing procedures for the determination of anti-HIV-1 IgG, anti-HBc IgG, anti-HTLV-I IgG, anti-HCV IgG, HBs antigen and anti-Treponema pallidum IgG.

Recall number
Z-0502-2026
Recalling firm
TECHNO-PATH MANUFACTURING LTD.
Product
Multichem ID-B. Model Number: SR103B. Intended for use as an unassayed positive qualitative quality control serum to monitor the precision of laboratory testing procedures for the determination of anti-HIV-1 IgG, anti-HBc IgG, anti-HTLV-I IgG, anti-HCV IgG, HBs antigen and anti-Treponema pallidum IgG.
Status
Open, Classified
Initiated
2025-10-17
Posted
2025-11-14
Terminated
not on file
Reason
Potential of negative recovery of the HBsAg analyte across two (2) lots of Multichem ID-B Positive Control when tested on the Alinity i platform.
Root cause
Process control
Action
LGC Clinical Diagnostics notified consignees on about 10/17/2025 via letter that Techno-path Manufacturing Ltd was recalling specific Multichem ID-B products. Consignee were informed that the Instructions for Use (IFU) for Multichem ID-B SR103B, Lots SB140524P and SB161024P have been updated to remove the representative reactivity for the analyte HBsAg; therefore, no representative reactivity result will be provided for HBsAg in these lots. Consignees were instructed to discard all previous versions of the IFU for affected lots, share the information with all applicable laboratory staff or customers if further distributed, retain the notification as part of the laboratory Quality System documentation, and complete and return the provided response form. Techno-path Manufacturing Ltd. is investigating this issue and will implement a correction for all future lots of Multichem ID-B Positive Control SR103B that will be released.
Quantity
51 units
Distribution
US distribution to CA, CO, CT, FL, HI, IL, LA, MI, MN, NV, NY, OR, SC, TN, TX, WI, WY.
Lot, serial or model codes
Model Number: SR103B. UDI: (01)05391523442312(240)SR103B(17)261031(10) SB161024P. Lot Number: SB161024P. Expiration Date: 31-10-2026.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OHQ
Firm FEI
3008156187
Firm city
Ballina Tipperary
Firm state
not on file
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06