Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories Ltd.: RANDOX Total Bilirubin.

Recall number
Z-0501-2024
Recalling firm
Randox Laboratories Ltd.
Product
RANDOX Total Bilirubin.
Status
Open, Classified
Initiated
2023-10-11
Posted
2023-12-06
Terminated
not on file
Reason
Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.
Root cause
Device Design
Action
Consignees were emailed an Urgent Device Correction letter on 10/11/2023. The notice instructed consignees to only use Total Bilirubin Vanadate Oxidation Method on Serum samples, to review results generated with Plasma samples in like with the clinical profiles of the patient, discuss the contents of the recall notice with your facility's Medical Director, and to return the provided response form completed to technical.services@randox.com within 5 working days. The recall notice is to be further distributed to all affected customers and those within consignee organizations, if further distributed. A second recall notification, dated 4/26/24, was sent to consignees identifying the reason for the product issue to be intralipid interference. Reagent IFUs have been updated with sample specific Plasma (lithium heparin) interference claims for intralipids, which can be found on www.randox.com. Consignees are asked to complete and return the provided response form to technical.services@randox.com within five working days.
Quantity
1644 units
Distribution
US Nationwide distribution in the states of AR, GA, NC, TN, & PR.
Lot, serial or model codes
Catalog No. BR8377, BR9766, & BR4061; GTIN: 05055273214772, 05055273208337, & 05055273214765; All Batches.
510(k) numbers
["K152344"]
PMA numbers
not on file
Product code
JFM
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register