Skip to the content

FDA Medical Device Recalls (openFDA)

Microgenics Corporation: MAS CardioImmune XL Cardiac Marker Control; Model: CAI-XL1; intended for use in the clinical laboratory as an assayed control serum for monitoring assay conditions in specific cardiac marker determinations.

Recall number
Z-0501-2015
Recalling firm
Microgenics Corporation
Product
MAS CardioImmune XL Cardiac Marker Control; Model: CAI-XL1; intended for use in the clinical laboratory as an assayed control serum for monitoring assay conditions in specific cardiac marker determinations.
Status
Terminated
Initiated
2014-10-20
Posted
2014-12-04
Terminated
2015-03-30
Reason
MAS CardioImmune XL, level 1 (lot CXL16011), is showing vial-to-vial variation for the analytes Creatine Kinase-MB (CK-MB) and B-Type Natriuretic Peptide-32 (BNP-32).
Root cause
Nonconforming Material/Component
Action
ThermoFisher sent an Urgent Medical Device Field Corrections dated October 13, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Foreign customers were sent letters via overnight mail and e-mail. ACTIONS TO BE TAKEN BY THE CUSTOMER: 1. Determine if you are using or have inventory of MAS Cardiolmmune XL, lot CXL16011. 2. Discontinue use and destroy any remaining inventory of MAS Cardiolmmune XL, lot CXL16011 per your local waste ordinances. 3. Retain a copy of this letter for your laboratory records. 4. If you have forwarded kits of MAS Cardiolmmune XL (lot CXL16011) to another laboratory, please provide a copy of this letter to them. 5. Complete the attached Medical Device Field Correction Response Form and return the form within 5 days to Thermo Fisher Scientific Technical Service as instructed in the form below. 6. Please call technical service at 1-800-232-3342, then press option 2, then option 2 or e-mail mas.controls@thermofisher.com to arrange immediate product replacement. If you are a distributor of the product, please contact your affected customer base, advise them of the situation, and provide them with a copy of this letter. You should insert your contact information, e-mail and fax numbers in the Medical Device Field Correction Response Form and request that they return the form to you. You should fill out the distributor section of the attached Medical Device Field Correction Response Form and return the form within 5 days to Thermo Fisher Scientific Technical Service as instructed in the form. Firm intends to try to get a response twice more by mail or telephone to insure we reach all customers. Affected product remaining at firm is quarantined and will be destroyed.
Quantity
2135 vials.
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AR, AZ, CA CO, CT, DE, FL GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WY., and the countries of Australia, Canada, China, Germany, Guam, Hong Kong and Taiwan.
Lot, serial or model codes
Model No.: CAI-XL1; Lot Number: CXL16011, Exp. Date: 2016-01-31.
510(k) numbers
["K061196"]
PMA numbers
not on file
Product code
JJY
Firm FEI
3010939897
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register