FDA Medical Device Recalls (openFDA)
Beaver-Visitec International Inc.: bvi Beaver EdgeAhead Safety Knife, (a) Sideport MVR .90mm 20G, REF 370531 (b) Sideport Angled, .60mm (23G), REF 378233 Product Usage: Specialty blades and knives are used for stab and cut incisions in ophthalmic, ENT and other surgical applications.
- Recall number
- Z-0500-2018
- Recalling firm
- Beaver-Visitec International Inc.
- Product
- bvi Beaver EdgeAhead Safety Knife, (a) Sideport MVR .90mm 20G, REF 370531 (b) Sideport Angled, .60mm (23G), REF 378233 Product Usage: Specialty blades and knives are used for stab and cut incisions in ophthalmic, ENT and other surgical applications.
- Status
- Terminated
- Initiated
- 2017-11-03
- Posted
- 2018-02-01
- Terminated
- 2023-05-31
- Reason
- The protective shield was retracted, and the blade was exposed.
- Root cause
- Process control
- Action
- The firm disseminated their safety advisory notices on 11/03/2017 by email and visit. The letter identified the affected product, problem and actions to be taken. The firm requested the following action:"Prior to opening the Tyvek tray, please visually confirm that the plastic shield is not retracted and the activation slider is fully engaged."
- Quantity
- 45, 680 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Italy, UK, Finland, Germany, Ireland, Turkey
- Lot, serial or model codes
- (a) REF 370531; Batch numbers: 3185248 3191981 3193837 3199247 3201734 3204737 3206382 3208096 3211720 3213361 3214546 3217371 3219431 3222626 3225976 (b) REF 378233; Batch numbers: 3191985 3204382 3225442
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- EMF
- Firm FEI
- 1211998
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28