Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: 1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.

Recall number
Z-0499-2022
Recalling firm
Siemens Medical Solutions USA, Inc
Product
1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.
Status
Open, Classified
Initiated
2021-11-23
Posted
2022-01-14
Terminated
not on file
Reason
Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.
Root cause
Software design
Action
On November 23, 2021, Siemens Healthineers issued an Urgent Medical Device Correction notice to all users of Sensis Vibe Hemo systems, Sensis and Sensis Vibe Combo systems with software version VD12A. Please make sure that an alternative system can be used to provide treatment. Please notify and instruct accordingly all the staff at your organization who need to be aware of this notice and will comply with the recommendations therein. Siemens will provide a software update to all affected customers which will correct the issue via Update Instruction AX077/21/S. Siemens service organization will contact you to arrange a date to perform this corrective action. Please feel free to contact Siemens service organization for an earlier appointment at 1-800-888-7436.
Quantity
1047
Distribution
US Nationwide distribution in the states of AK, AL, AR, AZ, CA, FL, GU, HI, IA, IL, IN, KS, KY, MA, MI, MO, MS, NC, NE, NJ, NY, OH, OK,PA, PR, SD, TN, TX, WA, WV.
Lot, serial or model codes
1. Sensis- 10764561; UDI: 04056869010137; Serial Numbers: 103493, 103432, 103466 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199; Serial Numbers: 103137, 103144, 103453, 103108, 103111, 103047, 103048, 103171, 103182, 103193, 103301, 103170, 103223,, 103225, 103247, 103313, 103315, 103196, 103198, 103166, 103173, 103177, 103184, 103186, 103199, 102326, 103139, 103161, 103167, 103168, 103172, 103178, 100263, 103424, 103319, 103320, 103321, 103323, 103325, 103328, 103330, 103333, 103334, 103335, 103344, 100216, 103337, 100416, 102008, 103413, 101083, 100208, 100209, 102138, 102145, 102147, 103160, 103314, 102199, 103079, 102221, 103123, 103124, 103131, 100163, 100306, 100309, 103300, 101099, 101100, 101101, 103265, 100422, 103019, 103331, 103362, 103393, 103402, 103419, 103252, 103261, 103267, 103269, 103270, 103276, 103277, 103280, 103281, 103022, 103381, 103194 103145, 103148, 103152, 100345, 102183, 103345, 103368, 103415, 103416, 103245, 103249, 103250, 103264, 103080, 103082, 103086, 103088, 103089, 103090, 100231, 103431, 103306, 103341, 103114, 100070, 100071, 100074, 102155, 102157, 103040, 103042, 103039, 102174, 102182, 103130, 103200, 103234, 103116, 103117, 103119, 103288, 103008, 103216, 103465, 103189, 103208, 102010, 102295, 102296, 102302, 100006, 100005, 103352, 103283, 103308, 103310, 103115, 103244, 103246, 103014, 103242, 102143, 103359, 103274, 103437 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205; Serial Numbers: 123021, 123004, 123047, 123053, 122011 ***Updated 1/18/22*** SENSIS Vibe Hemo, Serial #103185 ***Updated 2/24/22*** SENSIS Vibe Hemo Serial #103290 103510 103451 103461 103444 103450 103448 103445 103442 103449 103439 103457 SENSIS Vibe Combo Serial #123055 123056 123058 The following serial numbers have been removed from the scope of this recall: SENSIS Vibe Hemo Serial #103301 103424 ***Updated 8/5/22*** The following serial numbers have been added to the recall: SENSIS Vibe Combo Serial numbers 103307, 103312
510(k) numbers
["K150493"]
PMA numbers
not on file
Product code
DQK
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28