Covidien LLC dba Uni-Patch: Reusable, Self-Adhering TENS/NMES/FES electrodes Single Patient Use Only. An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.

Recall number
Z-0499-2013
Recalling firm
Covidien LLC dba Uni-Patch
Product
Reusable, Self-Adhering TENS/NMES/FES electrodes Single Patient Use Only. An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Status
Terminated
Initiated
2012-02-09
Posted
2012-12-10
Terminated
2013-01-03
Reason
Covidien is conducting a voluntary recall regarding specific lots of TENS electrodes because it was determined that during the manufacturing process the surface of the wires in these lots may have been exposed to an employee who had a minor cut.
Root cause
Employee error
Action
Covidien sent An Urgent Product Recall letter, dated February 9, 2012, to all affected consignees. The letter identified the affected product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory and quarantine the affected product if found. The letter asked customers to fill in teh quantity remaining in their inventory and fax it to 651-565-5888. A Covidien Customer Service rep would contact the consignee for returning product. Consignees were also requested to notify their customers of the recall. For questions regarding this recall call 508-261-8532.
Quantity
3594 pouches
Distribution
Nationwide Distribution including CA, IL, MA, MN, and SD.
Lot, serial or model codes
Lots 202404, 202506, 203431.
510(k) numbers
["K900519"]
PMA numbers
not on file
Product code
GXY
Firm FEI
1000116809
Firm city
Wabasha
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register