Skip to the content

FDA Medical Device Recalls (openFDA)

Alphatec Spine, Inc.: ILLICO Posterior Thoracolumbar Access Retractor, Part Number: 73520 The ILLICO Posterior Thoracolumbar Access Retractor is an instrument that is intended for use in a minimally invasive approach for surgical procedures. The retractor allows retraction of tissue through combination of articulating blades. Device will serve to provide access for surgeons to surgical site.

Recall number
Z-0498-2010
Recalling firm
Alphatec Spine, Inc.
Product
ILLICO Posterior Thoracolumbar Access Retractor, Part Number: 73520 The ILLICO Posterior Thoracolumbar Access Retractor is an instrument that is intended for use in a minimally invasive approach for surgical procedures. The retractor allows retraction of tissue through combination of articulating blades. Device will serve to provide access for surgeons to surgical site.
Status
Terminated
Initiated
2009-01-28
Posted
2009-12-11
Terminated
2009-12-22
Reason
The recall was initiated after Alphatec Spine received a complaint on the ILLICO Posterior Thoracolumbar access retractor. It was reported that one of the retractor arms in the retractor broke during surgery.
Root cause
Device Design
Action
On January 28, 2009 Alphatec Spine contacted their domestic consignees via telephone and the international consignee was contacted via telephone and electronic mail. Alphatec Spine's telephone call consisted of informing the consignees of the product affected, the problem, what actions needed to be taken and what action Alphatec Spine would take to arrange for pick-up of the affected products.
Quantity
14
Distribution
AZ, CA, OR, TX, and Tokyo, Japan.
Lot, serial or model codes
Lot number: 5271901
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAD
Firm FEI
2027467
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register