FDA Medical Device Recalls (openFDA)
Alphatec Spine, Inc.: ILLICO Posterior Thoracolumbar Access Retractor, Part Number: 73520 The ILLICO Posterior Thoracolumbar Access Retractor is an instrument that is intended for use in a minimally invasive approach for surgical procedures. The retractor allows retraction of tissue through combination of articulating blades. Device will serve to provide access for surgeons to surgical site.
- Recall number
- Z-0498-2010
- Recalling firm
- Alphatec Spine, Inc.
- Product
- ILLICO Posterior Thoracolumbar Access Retractor, Part Number: 73520 The ILLICO Posterior Thoracolumbar Access Retractor is an instrument that is intended for use in a minimally invasive approach for surgical procedures. The retractor allows retraction of tissue through combination of articulating blades. Device will serve to provide access for surgeons to surgical site.
- Status
- Terminated
- Initiated
- 2009-01-28
- Posted
- 2009-12-11
- Terminated
- 2009-12-22
- Reason
- The recall was initiated after Alphatec Spine received a complaint on the ILLICO Posterior Thoracolumbar access retractor. It was reported that one of the retractor arms in the retractor broke during surgery.
- Root cause
- Device Design
- Action
- On January 28, 2009 Alphatec Spine contacted their domestic consignees via telephone and the international consignee was contacted via telephone and electronic mail. Alphatec Spine's telephone call consisted of informing the consignees of the product affected, the problem, what actions needed to be taken and what action Alphatec Spine would take to arrange for pick-up of the affected products.
- Quantity
- 14
- Distribution
- AZ, CA, OR, TX, and Tokyo, Japan.
- Lot, serial or model codes
- Lot number: 5271901
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GAD
- Firm FEI
- 2027467
- Firm city
- Carlsbad
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28