Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: Graseby 3300 Pump and Graseby 3000 Series Pump Accessories Patient controlled syringe pump. Use for Infusion

Recall number
Z-0497-2018
Recalling firm
Smiths Medical ASD Inc.
Product
Graseby 3300 Pump and Graseby 3000 Series Pump Accessories Patient controlled syringe pump. Use for Infusion
Status
Terminated
Initiated
2017-08-24
Posted
not on file
Terminated
2020-05-01
Reason
In certain cases, the lock and key supplied with Graseby 3300 pumps and accessories does not match the standard Graseby 3300 key.
Root cause
Nonconforming Material/Component
Action
Notification was sent to customers on August 24, 2017. Included with the Product Safety & Recall Notification, Instructions to Customers and the Recall response Form. Customers who find product subject to the recall are to contact Smith Medical Customer Service Department to make arrangements to have the affected products returned to Smiths. For further questions, please call (763) 383-3072.
Quantity
77
Distribution
Worldwide Distribution to the countries of : Australia, Canada, Germany, France, Great Britain, Hong Kong, and Ireland
Lot, serial or model codes
Model: 0128-0001, 0128-0025, 0128-0135, 0128-0136, and 0128-0451 Lot/Serial #'s: 1074,8305,8718, 11325, 11540, 16230, 32982, 34076, 42678, 52734, 55337, 60959, 60965, 64104, 67638, 68079, 69199, 72740, 90734, 100817, 12001289, 12001409, 12001557, 12001612, 12008230, 004325, 008303, 012348, 018148, 032159, 041927, 048278, 054171, and 069850
510(k) numbers
["K915270"]
PMA numbers
not on file
Product code
MEA
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register