FDA Medical Device Recalls (openFDA)
Ev3, Inc.: TrailBlazer Support Catheter, ev3 inc, 4600 Nathan Lane North, Plymouth, MN 55442-2920, Rx Only, Sterile EO. US Model Numbers: SC-014-135, SC-018-090, SC-035-065, SC-035-135, SC-014-150, SC-018-150, SC-035-090, and SC-035-150. Intended to guide and support a guide wire during access of the vasculature.
- Recall number
- Z-0497-2010
- Recalling firm
- Ev3, Inc.
- Product
- TrailBlazer Support Catheter, ev3 inc, 4600 Nathan Lane North, Plymouth, MN 55442-2920, Rx Only, Sterile EO. US Model Numbers: SC-014-135, SC-018-090, SC-035-065, SC-035-135, SC-014-150, SC-018-150, SC-035-090, and SC-035-150. Intended to guide and support a guide wire during access of the vasculature.
- Status
- Terminated
- Initiated
- 2009-11-06
- Posted
- 2009-12-23
- Terminated
- 2011-01-05
- Reason
- Catheter may crack near the radiopaque marker band at the distal end of the device during use. Cracking of the catheter shaft in this location can result in material separation, and potential embolization.
- Root cause
- Component design/selection
- Action
- Consignees were sent a Medical Device Recall letter, dated 11/6/09. The letter was addressed to Risk Manger or Cath Lab Manage. The letter provided the Issue Summary, identified the Affected Product, and stated the Required Action, which asked consignees to locate and remove the affected product. Questions should be directed to ev3 Customer Service at 800-716-6700.
- Quantity
- 70
- Distribution
- Nationwide Distribution -- USA, including states of CA, FL, GA, IL, IN, IA, KS, KY, MI, MT, NJ, PA, SD, TN, TX, and WI.
- Lot, serial or model codes
- Lot #: 7828282, 7792290, 7792584, 7805570, 7805797, 7806392, 7820252, 7790666, 7791887, 7803307, 7835331, 7822400, 7822593, 7800446, 7800555, 7800756, 7800779, 7800809, 7801822, 7801875, 7803305, 7803306, 7820273, 7834779, 7834845, 7824905, and 7832205.
- 510(k) numbers
- ["K092299"]
- PMA numbers
- not on file
- Product code
- DQY
- Firm FEI
- 2183870
- Firm city
- Plymouth
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28