FDA Medical Device Recalls (openFDA)
Applied Medical Technology Inc: Applied Medical Technology, Inc., NutraGlide, 5 F- 6 F x 90 Nasal Feeding Tube with Stylet and ENfit, REF NFT-05090-I
- Recall number
- Z-0496-2026
- Recalling firm
- Applied Medical Technology Inc
- Product
- Applied Medical Technology, Inc., NutraGlide, 5 F- 6 F x 90 Nasal Feeding Tube with Stylet and ENfit, REF NFT-05090-I
- Status
- Completed
- Initiated
- 2025-10-06
- Posted
- 2025-11-13
- Terminated
- not on file
- Reason
- The nasal feeding tube may be occluded
- Root cause
- Process control
- Action
- Applied Medical Technology, Inc. issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 10/06/2025 via certified mail. The notice explained the issue, potential risk to health, and requested the consignee quarantine the affected product, complete and return the acknowledgement, and contact Applied Medical Technology, Inc. either by telephone at 440-717-4000 or via email at cs@appliedmedical.net to arrange for the return of the affected product. Those who further distributed the product were directed to notify those to whom the product was transferred of this recall. For questions or concerns, contact: Customer Service, 440-717-4000 (tel), 440-717-4200 (fax), cs@appliedmedical.net (email). Hours of operation 8:30 A - 5:00 P, Eastern time
- Quantity
- not on file
- Distribution
- US Nationwide distribution in the states of MA, RI, NC, OH.
- Lot, serial or model codes
- Box P/N NFT-05090-I, Lot Number 250821-238, exp. 2027-10-01, UDI (01)00842071149666(17)271001(10)250821-238, Pouch P/N ENFT-05090-I, Lot Number 250812-470, exp. 2027-11-01, UDI (01)00842071149543(17)271101(10)250812-470; Box P/N NFT-05090-I, Lot Number 250815-355, exp. 2027-10-01, UDI (01)00842071149666(17)271001(10)250815-355, Pouch P/N ENFT-05090-I, Lot Number 250812-468, exp. 2027-10-01, UDI (01)00842071149543(17)271001(10)250812-468.
- 510(k) numbers
- ["K202539"]
- PMA numbers
- not on file
- Product code
- KNT
- Firm FEI
- 1526012
- Firm city
- Brecksville
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28