Skip to the content

FDA Medical Device Recalls (openFDA)

Cure Medical LLC: Pediatric, 10Fr, Intermittent Urinary Catheters Model: P10 Pouch Label UPC (P10) *+M405P100C* Intermediate Box Label UPC (P10) *+M405P103F* Carton Label UPC (P10) *+M405P105H* The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches). There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case. Intended to be used by pediatric males and females for the purpose of bladder drainage

Recall number
Z-0496-2015
Recalling firm
Cure Medical LLC
Product
Pediatric, 10Fr, Intermittent Urinary Catheters Model: P10 Pouch Label UPC (P10) *+M405P100C* Intermediate Box Label UPC (P10) *+M405P103F* Carton Label UPC (P10) *+M405P105H* The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches). There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case. Intended to be used by pediatric males and females for the purpose of bladder drainage
Status
Terminated
Initiated
2014-11-11
Posted
2014-12-16
Terminated
2015-04-10
Reason
Some of the intermediate boxes within lot 140417-3, were labeled as P8, a French Size 8 pediatric catheter but actually contain P10, French Size 10 pediatric catheters. Trying to use a P10 catheter for bladder drainage, where P8 catheters have been prescribed, may result in acute urinary retention; and, forced attempts to insert the French Size 10 catheter lead to urethral trauma.
Root cause
Labeling mix-ups
Action
Cure Medical sent Urgent: Medical Device Correction and Removal Notification, dated November 11, 2014, to customers. The letter identified the affected product and the reason for the recall. Customers were asked to immediately quarantine all cases of P10 catheters, with Lot Number 140417-3, and destroy any boxes that are labeled as P8, Lot 140410-3 and contact us for replacement products. Additionally, if any cases of P10, Lot 140417-3 products or boxes of P8 products, Lot Number 140410-3 were further distributed, those customers were to be identified and notified of the problem via the enclosed letter, which includes instructions on what to do with any affected product. Cure Medical, LLC will work in cooperation with customers to cover expenses associated with this action. Customers were to complete and return the enclosed response form as soon as possible. If there are any questions, customers are to call Carole Harris at 949-723-0364, Monday-Friday, between 09:30 am and 05:00pm Pacific Time.
Quantity
192 cases = 1,920 boxes = 57,600 each
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot Number: 140417-3
510(k) numbers
["K110653"]
PMA numbers
not on file
Product code
EZD
Firm FEI
3005471919
Firm city
Newport Beach
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register