Medtronic Neuromodulation: SynchroMed EL, models 8626-10, 8626L-10, 8626-18, 8626L-18, 8627-10, 8627L-10, 8627-18, 8627L-18. implantable infusion pump. The pump is part of an implanted infusion system that consists of a SynchroMed EL pump and a catheter. The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The SynchroMed EL Infusion System with a 10 mL reservoir volume is indicated for use in the smaller patient who has insufficient body mass to support the larger SynchroMed EL pump or in patients who choose a smaller, lower profile pump when presented with the choice, when patient therapy requires the chronic infusion of drugs or fluids referred to in this manual.
- Recall number
- Z-0496-2013
- Recalling firm
- Medtronic Neuromodulation
- Product
- SynchroMed EL, models 8626-10, 8626L-10, 8626-18, 8626L-18, 8627-10, 8627L-10, 8627-18, 8627L-18. implantable infusion pump. The pump is part of an implanted infusion system that consists of a SynchroMed EL pump and a catheter. The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The SynchroMed EL Infusion System with a 10 mL reservoir volume is indicated for use in the smaller patient who has insufficient body mass to support the larger SynchroMed EL pump or in patients who choose a smaller, lower profile pump when presented with the choice, when patient therapy requires the chronic infusion of drugs or fluids referred to in this manual.
- Status
- Terminated
- Initiated
- 2012-11-09
- Posted
- 2012-12-13
- Terminated
- 2017-12-31
- Reason
- Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.
- Root cause
- Nonconforming Material/Component
- Action
- The firm Medtronic, sent an "Urgent medical Device Safety Notification" dated November 2012, to its customers beginning November 09, 2012. The letter described the product, problem and actions to be taken. The customers were instructed to follow the recommendations provided in the letter and complete and return the PHYSICIAN REPLY FORM via mail at Medtronic Neuromodulation Quality Assurance, 7000 Central Ave NE, Mailstop RCC150, Minneapolis, MN 55432; fax to: 800-897-3899 or email to: neuro.quality@medtronic.com. If you have questions, please contact Medtronic Neuromodulation Technical Services at 1-800-707-0933 weekdays 7am - 6pm CST. UPDATE 2/1/2016: Medtronic began notifying consignees 2/1/2016 via, mail, telephone and/or in person visit by a Medtronic Field Representative. The Medtronic Field Representatives will retrieve devices in the US and provide an Urgent Medical Device Removal letter. The letter states that device design has been updated and the Medtronic Field Representative has retrieved any devices with a Use By date on or before 5/14/2017. An account specific Customer Confirmation Form was to be completed by either the Medtronic Field Representative or the consignee. Consignees with questions can contatct their Medtronic Field Representative, or Customer Service at 1-888-854-0978.
- Quantity
- 174,700 devices total
- Distribution
- Worldwide distribution: USA (nationwide) and countries including: Canada and Mexico.
- Lot, serial or model codes
- All SynchroMed EL and SynchroMed II pumps
- 510(k) numbers
- not on file
- PMA numbers
- ["P860004S042"]
- Product code
- LKK
- Firm FEI
- 2182207
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28