Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Syngo.plaza Picture Archiving and Communication System (PACS) with software version VB20A; Model numbers: 10863171, 10863172, 10863173 Syngo.plaza is a Picture Archiving and Communication System (PACS) software device intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.

Recall number
Z-0495-2018
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Syngo.plaza Picture Archiving and Communication System (PACS) with software version VB20A; Model numbers: 10863171, 10863172, 10863173 Syngo.plaza is a Picture Archiving and Communication System (PACS) software device intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
Status
Terminated
Initiated
2017-08-18
Posted
not on file
Terminated
2018-08-30
Reason
Software update to correct several issues that include (1)Potential data loss, (2) Study mix-up, (3) Incorrect measurements on multi-frame images, (4) Dearchiving issue, and (5) Unauthorized access of data due to inadequate permissions for shared folders.
Root cause
Software design
Action
Siemens released a software update VB20A_HF04 and a Customer Information Letter was sent with Update Instruction SY010/17/P. Service reports for the affected sites will be monitored to ensure the recall was effective. For further questions, please call (610) 219-6461.
Quantity
47
Distribution
USA (nationwide) Distribution to the states of : AL, CT, FL, KS, LA, MA, MI, MN, MO, MS, NE, NY, OH, OR, PA, TX and WI. Foreign consignees were not reported as device is manufactured outside the U.S.
Lot, serial or model codes
Syngo.plaza with software version VB20A; Model numbers: 10863171, 10863172, 10863173
510(k) numbers
["K132532"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register