FDA Medical Device Recalls (openFDA)
Zimmer CAS: PSI Knee System- Prosthesis, knee patellofemorotibial, semi-constrained, cemented, polymer/meal/polymer The Zimmer PSI Knee System is indicated as an orthopedic instrument system to assist in the positioning of knee replacement components.
- Recall number
- Z-0495-2015
- Recalling firm
- Zimmer CAS
- Product
- PSI Knee System- Prosthesis, knee patellofemorotibial, semi-constrained, cemented, polymer/meal/polymer The Zimmer PSI Knee System is indicated as an orthopedic instrument system to assist in the positioning of knee replacement components.
- Status
- Terminated
- Initiated
- 2014-11-03
- Posted
- 2014-12-16
- Terminated
- 2015-02-13
- Reason
- Internal discovery found that the MRI scans in some PSI knee guides involved an additional rotation adjustments that could potentially lead to misalignment due to a design flaw.
- Root cause
- Software design
- Action
- The firm, Zimmer CAS, sent an "URGENT: MEDICAL DEVICE NOTIFICATION" dated November 3, 2014 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to read this notice and be familiar with its contents; complete and return the attachment 1 CERTIFICATE OF ACKNOWLEDGEMENT form via fax to Zimmer Inc. at (574) 372-4265 or scan and email to corporatequality.postmarket@zimmer.com; and maintain a copy of the completed attachment 1 for your files. If you have questions concerning this notice, please call 514-861-4074 ext 292 or email: jean-sebastien.merette@zimmercas.com
- Quantity
- 15
- Distribution
- US Distribution in states of: AZ, CO, IL, OR, and WA.
- Lot, serial or model codes
- BKL0398L 20-8070-002-01; DHA0645L 20-8070-002-01; DHA0646R 20-8070-002-02; ECO0516L 20-8070-002-01; GHE0651R 20-8070-002-02; KAD0595L 20-8070-002-01; KWI0523R 20-8070-002-02; MHA0483L 20-8070-002-01; MHA0484R 20-8070-002-02; MKE0659R 20-8070-002-02; MMA0728R 20-8070-002-02; SSC0782L 20-8070-001-01; TCA0582L 20-8070-002-01; THO0803L 20-8070-002-01; TMA0760L 20-8070-002-01; Expiration dates 5/8/14-12/31/15
- 510(k) numbers
- ["K131409"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 3003714383
- Firm city
- Montreal
- Firm state
- not on file
- Firm country
- Canada
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28