Philips Healthcare Inc.: Philips IntelliVue Information Center/Client and Philips IntelliVue Information Center iX The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors.

Recall number
Z-0495-2013
Recalling firm
Philips Healthcare Inc.
Product
Philips IntelliVue Information Center/Client and Philips IntelliVue Information Center iX The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors.
Status
Terminated
Initiated
2012-11-09
Posted
2012-12-10
Terminated
not on file
Reason
Internal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms.
Root cause
Component design/selection
Action
Philips sent an URGENT - Medical Device Correction letter dated November 9, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Philips is conducting a voluntary correction to replace the audio cable on affected devices. Customer were instructed to follow the "Action to be taken by CustomerUser" section of the instructions. Customers with wquestions were instructed to contact their Phililps representataive or contact the Customer Care Solutions Center at (800) 722-9377. For questions regarding this recall call 978-687-1501.
Quantity
3165
Distribution
Worldwide Distribution - USA (nationwide) and Canada.
Lot, serial or model codes
PIIC 865435 M3140 Information Center Low Acuity 865436 M3150 Information Center Local DB 865417 M3145 Information Center Small NW DB 865418 M3155 Information Center NW DB 865415 M3151 Information Center Client 865421 M3170 Patient Link 865416 M3177 Trend Display 865422 M3140 Upgrade 865423 M3150 Upgrade 865426 M3145 Upgrade 865427 M3155 Upgrade 865424 M3151 Upgrade 865266 - M3140 Information Center Low Acuity 865267 M3150 Information Center Local DB 865269 M3145 Information Center Small NW DB 865268 M3155 Information Center NW DB 865270 M3151 Information Center Client 865274 M3170 Patient Link 865271 M3177 Trend Display 865277 M3140 Upgrade 865278 M3150 Upgrade 865281 M3145 Upgrade 865282 M3155 Upgrade 865279 M3151 Upgrade  PIIC iX 866023 IntelliVue Info Center iX 866117 PIIC Classic Upgrade 866027 PIIC iX HW Upgrade
510(k) numbers
["K081983","K102495"]
PMA numbers
not on file
Product code
MHX
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register