Skip to the content

FDA Medical Device Recalls (openFDA)

AMD Medicom Inc.: MEDICOM SAFEMASK PREMIER PLUS ASTM LEVEL 2 MASKS, Item Code 2040

Recall number
Z-0494-2020
Recalling firm
AMD Medicom Inc.
Product
MEDICOM SAFEMASK PREMIER PLUS ASTM LEVEL 2 MASKS, Item Code 2040
Status
Terminated
Initiated
2019-10-11
Posted
not on file
Terminated
2021-09-14
Reason
Due to an error on the packaging line, some level 2 masks may have been packed in some level 3 mask boxes. The two masks are visually indistinguishable, causing the risk of using a mask with a lower protection level than intended.
Root cause
Under Investigation by firm
Action
Urgent Medical Device Recall notification letters dated 10/11/19 were sent to customers. AMD Medicom requests that you discontinue distribution/use of and return all products from the catalog/lot number referenced above, for a credit. Our records indicate that you have purchased one or more of product code 2040 for re-sale to your customers. You should complete the attached Acknowledgment Form for inventory you have received and/or which is still in stock. AMD Medicom requests that you complete and return the Acknowledgment Form enclosed with this letter, even if you do not have the above-mentioned lot numbers in your inventory, within three (3) working days. In addition, please contact your affected customer base, advise them of the recall and provide them with a copy of this letter. You should insert your contact information, email and fax numbers in the Acknowledgement Form and request that they return the Acknowledgment Form to you. Please have them return any unused inventory to you for reconciliation. Once you have received all inventory from your customers, please contact Medicom customer service at 1-800-308-6589 or claims@medicom.ca to receive a Sales Return Order number to identify your shipment of returns. Please return all units of the affected lots (unsold inventory and returns) to: AMO Medicom Inc. 6054 Shook Road, Suite 200 Lockbourne, OH USA 43137 ATTN: Recall You may use the following courier information: UPS Account# V28W12
Quantity
756 cases (500 units/case)
Distribution
The products were distributed tot he following US states: CA, FL, IA, IN, MD, MI, NV, NY, PA, SC, TN, TX, and WA. The products were distributed to the following foreign countries: Canada.
Lot, serial or model codes
Lot 1917
510(k) numbers
["K051291"]
PMA numbers
not on file
Product code
FXX
Firm FEI
3002677549
Firm city
Lachine
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register