FDA Medical Device Recalls (openFDA)
Aircraft Medical Limited: McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items, each individually peel-pouched and EtO sterilized. Product Usage: Single Use, sterile video laryngoscope blade for use with McGRATH MAC Video Laryngoscopes
- Recall number
- Z-0494-2015
- Recalling firm
- Aircraft Medical Limited
- Product
- McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items, each individually peel-pouched and EtO sterilized. Product Usage: Single Use, sterile video laryngoscope blade for use with McGRATH MAC Video Laryngoscopes
- Status
- Terminated
- Initiated
- 2014-11-04
- Posted
- 2014-12-02
- Terminated
- 2017-11-17
- Reason
- The blade retaining clip may be difficult to engage and thereby increase the possibility of a blade becoming disconnected during intubation, specifically when the device is being retracted from the patients mouth.
- Root cause
- Component change control
- Action
- Aircraft Medical Limited sent a recall notification letter dated November 4, 2014 to the sole US consignee. The letter included instructions for the consignee to: 1) check its own stock and its customers' stock for the recalled products; 2) arrange for any remaining stock to be removed from use. Customers with any questions can contact Aircraft Medical by e-mail at ac@aircraftmedical.com.
- Quantity
- 40,900 blades (US); 16,000 blades (Foreign)
- Distribution
- Worldwide Distribution - US Nationwide in the state of GA and the countries of Austria, China, Denmark, Finland, France, Hong Kong, Israel, Italy, Netherlands, Poland, Saudi Arabia, Singapore, South Korea, Taiwan, and Turkey.
- Lot, serial or model codes
- Product Code: 350-017-000. Lot Numbers and Expiration Dates: Lot 13121202, Expires 2016/12/11; Lot 13122301, Expires 2016/12/25; Lot 13042401, Expires 2017/5/5; Lot 14030501, Expires 2017/3/16; Lot 14031814, Expires 2017/3/26; Lot 14061616, Expires 2017/6/25.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- CCW
- Firm FEI
- 3005416399
- Firm city
- EDINBURGH
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28