FDA Medical Device Recalls (openFDA)

Tosoh Bioscience Inc: TosohAutomated Enzyme Immunoassay Analyzer AIA-900, Product Catalog No. 022930. Product Usage: The AIA-900 Automated Enzyme Immunoassay System is an automated analyzer for in vitro diagnostic use that measures a variety of substances in serum, plasma, and or urine using the AIA-PACK reagent series.

Recall number
Z-0491-2018
Recalling firm
Tosoh Bioscience Inc
Product
TosohAutomated Enzyme Immunoassay Analyzer AIA-900, Product Catalog No. 022930. Product Usage: The AIA-900 Automated Enzyme Immunoassay System is an automated analyzer for in vitro diagnostic use that measures a variety of substances in serum, plasma, and or urine using the AIA-PACK reagent series.
Status
Terminated
Initiated
2017-06-11
Posted
not on file
Terminated
2019-12-31
Reason
This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.
Root cause
Nonconforming Material/Component
Action
Customers were sent a recall notification by email on July 11, 2017 informing them that their local Field Service Engineer would contact them to make arrangements to proactively replace the affected parts. Follow up notifications will be by Federal Express with verification of delivery.
Quantity
565 total devices U.S.
Distribution
Worldwide Distribution - U.S. Nationwide - Latin America
Lot, serial or model codes
All Photo Interrupter sensors that are current in AIA-900 analyzers with serial number(S/N) 10000103 to S/N 11147003.
510(k) numbers
not on file
PMA numbers
not on file
Product code
KHO
Firm FEI
3005529799
Firm city
Grove City
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register