Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine

Recall number
Z-0489-2025
Recalling firm
Roche Diagnostics Operations, Inc.
Product
cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine
Status
Open, Classified
Initiated
2024-10-17
Posted
2024-11-19
Terminated
not on file
Reason
Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)
Root cause
Under Investigation by firm
Action
Roche issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 10/17/2024 via UPS Ground. The notice explained the issue, clinical significance, and requested the following actions be taken: "" HCYS kits lot number 747029 expired on September 30, 2024. Ensure any inventory you have of this lot has been discarded per your local guidelines. " Follow your laboratory s standard operating procedures to determine specific clinical implications for your patients and whether retrospective review of patient results is required. " Consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site." For questions: contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336
Quantity
458 units
Distribution
US: NJ, TX, NY, FL, IN, TX, OK, SC, IA,
Lot, serial or model codes
UDI/DI 04015630929979, Lot Number 747029, exp. 30-SEPT-2024
510(k) numbers
["K113793"]
PMA numbers
not on file
Product code
LPS
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register