Skip to the content

FDA Medical Device Recalls (openFDA)

Philips North America, LLC: MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician

Recall number
Z-0489-2018
Recalling firm
Philips North America, LLC
Product
MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Status
Terminated
Initiated
2017-07-12
Posted
2018-02-08
Terminated
2020-05-27
Reason
Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
Root cause
Nonconforming Material/Component
Action
Philips Healthcare sent an Urgent Medical Device Correction Notification/Field Safety Notice letter dated June, 2017 to their customers. The letter identified the affected product, problem, hazard involved and actions to be taken. Philips is requesting that customers report if their device fails to function as intended. Philips will contact all customers for device replacement once parts are available. If you need further information or support concerning this action, please contact your local Philips representative.
Quantity
71 units
Distribution
No U.S. distribution, Internationally to: India, Ireland, Japan, Republic of Korea, Malaysia, Mexico, South Africa & United Kingdom
Lot, serial or model codes
Serial Numbers:  US00601802, US00601968, US00601969, US00602114, US00602122, US00602124, US00602125, US00602126, US00602127, US00602128, US00602145, US00602180, US00602182, US00602184, US00602185, US00602186, US00602187, US00602189, US00602206, US00602236, US00602237, US00602238, US00602239, US00602240, US00602241, US00602242, US00602243, US00602244, US00602245, US00602246, US00602247, US00602248, US00602250, US00602251, US00602252, US00602253, US00602254, US00602344, US00602345, US00602346, US00602347, US00602348, US00602349, US00602350, US00602351, US00602352, US00602353, US00602354, US00602355, US00602356, US00602357, US00602358, US00602359, US00602360, US00602361, US00602362, US00602363, US00602364, US00602365, US00602366, US00602367, US00602368, US00602369, US00602370, US00602371, US00602372, US00602373, US00602374, US00602375, US00602376, US00602377
510(k) numbers
["K031187","K040404","K051134","K061707","K062233"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register