FDA Medical Device Recalls (openFDA)
Biomet Microfixation, Inc.: LACTOSORB SYSTEM 2.0MM ADJUSTABLE SELF-DRILLING TAP, REF: 915-2075, NON-STERILE PRODUCT, BIOMET MICROFIXATION,K INC. 1520 TRA DEPORT DRIVE JACKSONVILLE, FL 32218 USA. www.biometmicrofixation.com. Lacto Sorb Self Drilling Taps are used to drill and tap holes. They are self drilling so they do not require a hole to be drilled prior to use. They are used along with the Lacto SorbBulbs Handles or Power Drivers to advance the Tap into the bone.
- Recall number
- Z-0489-2010
- Recalling firm
- Biomet Microfixation, Inc.
- Product
- LACTOSORB SYSTEM 2.0MM ADJUSTABLE SELF-DRILLING TAP, REF: 915-2075, NON-STERILE PRODUCT, BIOMET MICROFIXATION,K INC. 1520 TRA DEPORT DRIVE JACKSONVILLE, FL 32218 USA. www.biometmicrofixation.com. Lacto Sorb Self Drilling Taps are used to drill and tap holes. They are self drilling so they do not require a hole to be drilled prior to use. They are used along with the Lacto SorbBulbs Handles or Power Drivers to advance the Tap into the bone.
- Status
- Terminated
- Initiated
- 2009-09-15
- Posted
- 2010-01-04
- Terminated
- 2010-01-23
- Reason
- Biomet Microfixation is recalling LactoSorb Self Drilling Adjustable Tap Model Numbers 915-2075 Lot # 064350 and 915-2185 Lot # 885580. The two lots identified have a new style stop and an old style tap; they are not compatible, as the depth etchings do not line up correctly. Depending on the tap (1.5mm VS 2.0mm) you may drill too shallow or too deep.
- Root cause
- Component change control
- Action
- Biomet Microfixation notified consignees of the recalled product beginning September 2009 via telephone and e-mail correspondence. Users were asked to return the affected product to the firm. For further information, contact Biomet Microfixation at 1-904-741-9468.
- Quantity
- 15 pieces
- Distribution
- USA - FL, MO, and KS; Foreign: Brazil and China.
- Lot, serial or model codes
- Lot Number: 064350
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HWC
- Firm FEI
- 1032347
- Firm city
- Jacksonville
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28