Skip to the content

FDA Medical Device Recalls (openFDA)

TMJ Solutions Inc: TMJ Bilateral Implant REF: CHG020 Sterile EO, Rx Only

Recall number
Z-0488-2026
Recalling firm
TMJ Solutions Inc
Product
TMJ Bilateral Implant REF: CHG020 Sterile EO, Rx Only
Status
Open, Classified
Initiated
2025-10-13
Posted
2025-11-14
Terminated
not on file
Reason
Patient-fitted temporomandibular joint implant contained discrepancy in screw hole position.
Root cause
Process control
Action
On February 25, 2026 Stryker (TMJ Solutions) issued a " Urgent: Medical Device Recall" Expansion notification to additional, affected consignees. Initial notification and previous expansion was issued on September 30, 2025 and December 4, 2025. TMJ Solutions asked consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product (Attachment B). Segregate the affected products, quarantine and discontinue use. 3. Please complete the Business Reply Form or use the QR Code on Attachment A. Response is required. a. If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility .
Quantity
63 units
Distribution
US: IL, OR, MS, TX, OK, UT, MA, MD, NY, PA, VA, FL, CA, MN, GA OUS: Canada, Spain, Brazil,
Lot, serial or model codes
UDI: 07613327626551,07613327626568,07613327626575,07613327626605,07613327627749 07613327627848/ Lot #s: 2410301062 2306091054 2312121078 2401181023 2401231058 2403191046 2404291064 2405021079 2407221023 2407241032 2408081047 2408131037 2408161043 2409031033 2409241016 2410011097 2410071065 2410091014 2410111030 2410291062 2411111012 2411121044 2411201015 2412061031 2412121062 2412201009 2501021011
510(k) numbers
not on file
PMA numbers
["P980052"]
Product code
LZD
Firm FEI
3001237463
Firm city
Ventura
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06