Philips Medical Systems: HeartStart FR2+ AED with Text Display, Model 989803148611. This product is G2005 Update in accordance with the American Heart Association 2005 Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care (ECC). The label of the device states: "HEARTSTART...FR2+ Automated External Defibrillator...Manufactured by: Philips Medical Systems Seattle, WA 98121".

Recall number
Z-0488-2011
Recalling firm
Philips Medical Systems
Product
HeartStart FR2+ AED with Text Display, Model 989803148611. This product is G2005 Update in accordance with the American Heart Association 2005 Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care (ECC). The label of the device states: "HEARTSTART...FR2+ Automated External Defibrillator...Manufactured by: Philips Medical Systems Seattle, WA 98121".
Status
Terminated
Initiated
2010-09-22
Posted
2010-11-29
Terminated
2013-09-05
Reason
A flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture.
Root cause
Nonconforming Material/Component
Action
On 10/5/10, Philips began sending out the Voluntary Medical Device Removal notification letter to their consignees. This letter informs the consignees that Philips Healthcare is conducting a voluntary product removal of a limited number of HeartStartFR2+ automated external defibrillators (AEDs), model numbers M3860A, M3861A, M3849A, and M3841A, manufactured by Philips and shipped between March 2007 and August 2010. The owners of affected AEDs are advised to run a battery insertion test. If it passes, the AED should remain in service until its replacement is received. Philips will be replacing affected units. The consignees can visit the firm's website www.philips.com/FR2AEDAction or contact Philips at 1-800-263-3342 for any information or support concerning the issue.
Quantity
6 units
Distribution
Devices were distributed to hospitals, medical facilities, and distributors through out the US and to foreign consignees. Foreign consignees including countries: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Ecuador, El Salvador, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Guatemala, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Japan, Jordan, Kampuchea, Kenya, Korea, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Malta & Gozo, Mexico, Netherland, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestinian Territory, Panama, Peru, Philippines, Portugal, Puerto Rico, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, United Arab Emirates, United Kingdom, Vietnam, and Zimbabwe.
Lot, serial or model codes
Serial Numbers:   0205001035, 0307230040, 0307230281, 1204000353, 1204000405, and 1204000766.
510(k) numbers
["K003565","K013425"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1000524572
Firm city
Seattle
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register