FDA Medical Device Recalls (openFDA)

Boston Scientific Corp: Boston Scientific Easy Core Biopsy Device, 18 ga. x 21 cm, sterile, latex free. Boston Scientific, Spencer, IN; UPN 500-104, REF M0065001040 and M0065001041. For use to endoscopically or percutaneously retrieve tissue samples of soft organs, tumors or masses for histological analysis. Soft tissue sampling includes, but is not limited to, organs such as breast, liver, kidney or prostate.

Recall number
Z-0487-2009
Recalling firm
Boston Scientific Corp
Product
Boston Scientific Easy Core Biopsy Device, 18 ga. x 21 cm, sterile, latex free. Boston Scientific, Spencer, IN; UPN 500-104, REF M0065001040 and M0065001041. For use to endoscopically or percutaneously retrieve tissue samples of soft organs, tumors or masses for histological analysis. Soft tissue sampling includes, but is not limited to, organs such as breast, liver, kidney or prostate.
Status
Terminated
Initiated
2008-10-17
Posted
2009-01-12
Terminated
2009-12-15
Reason
These devices may experience difficulty cocking and arming, resulting in an inability to use the devices.
Root cause
Nonconforming Material/Component
Action
U.S. consignees were notified via letter dated 10/17/08, which informs them of the problem, instructs them to remove from use and quarantine the recalled lots and requests completion and submission of the Reply Verification Tracking form via fax at 508-683-5578. For questions about this recall, contact your local Boston Scientific representative.
Quantity
14,254 of all devices.
Distribution
Worldwide Distribution including USA and countries of Algeria, Austria, Australia, Belgium, Canada, Denmark, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Ireland, Italy, Japan, Mexico, Netherlands, Norway, Philippines, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Japan and United Arab Emirates.
Lot, serial or model codes
Lots 0011520534, 0011539483 and 0011543121.
510(k) numbers
["K040893"]
PMA numbers
not on file
Product code
FCG
Firm FEI
1828132
Firm city
Spencer
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register