Skip to the content

FDA Medical Device Recalls (openFDA)

Ascom (US) Inc.: Ascom Telligence Nurse Call System

Recall number
Z-0486-2019
Recalling firm
Ascom (US) Inc.
Product
Ascom Telligence Nurse Call System
Status
Open, Classified
Initiated
2018-11-02
Posted
not on file
Terminated
not on file
Reason
There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.
Root cause
Software Manufacturing/Software Deployment
Action
Ascom Partners / Distributors were contacted by e-Mail, USPS Mail and Phone as necessary on 11/02/2018 to determine where the Telligence 5.0 Product is installed.
Quantity
57 units
Distribution
AL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX UT, VA, WI
Lot, serial or model codes
Version 5.0 (Build 5.0.4)
510(k) numbers
not on file
PMA numbers
not on file
Product code
IQA
Firm FEI
3005334549
Firm city
Lakewood Ranch
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register