FDA Medical Device Recalls (openFDA)

Abbott Diabetes Care, Inc.: FreeStyle Lite Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count; 100 count Product of Ireland; UPC 6 99073 70819 9; - 50 count UPC 6 99073 70822 9 - 50 count UPC 6 99073 71026 0 - 50 count UPC 6 99073 70827 4 - 100 count Manufactured by Abbott Diabetes Care Inc. Alameda, CA. The FreeStyle Blood Glucose test strip is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger, upper arm and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.

Recall number
Z-0486-2014
Recalling firm
Abbott Diabetes Care, Inc.
Product
FreeStyle Lite Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count; 100 count Product of Ireland; UPC 6 99073 70819 9; - 50 count UPC 6 99073 70822 9 - 50 count UPC 6 99073 71026 0 - 50 count UPC 6 99073 70827 4 - 100 count Manufactured by Abbott Diabetes Care Inc. Alameda, CA. The FreeStyle Blood Glucose test strip is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger, upper arm and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.
Status
Terminated
Initiated
2013-11-18
Posted
2013-12-17
Terminated
2015-08-03
Reason
Certain lots of FreeStyle and FreeStyle Lite Blood Glucose Test Strips produce erroneously low blood glucose results when using FreeStyle Blood Glucose Meters, FreeStyle Flash Blood Glucose Meters and the FreeStyle blood glucose meter built into the OmniPod system. Erroneously low results that are not recognized may pose significant risks to your health.
Root cause
Other
Action
Abbott Diabetes Care sent an Urgent Product Recall letter in November 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers who were affected were instructed to take the following steps: Call Abbott Diabetes Care Customers Service immediately at 1-888-736-9869 for replacement of the affected product. While waiting for replacement strips to arrive, use an alternate method to measure their blood glucose. If only test strips available to customers were from affected lots, customers should not stop testing their blood glucose. Customers were instructed to dispose of the test strips in accordance with their local and state disposal regulations. Customers with questions were instructed to call Abbott Diabetes Care Customers Service at 1-888-736-9869. For questions regarding this recall call 510-239-2775. Abbott issued a press release at the request of CDRH on november 27, 2013.
Quantity
141,219 cartons for human use - total
Distribution
Worldwide Distribution - USA (nationwide and Puerto Rico) and Internationally to Australia, Austria, Belgium, Denmark, Egypt, Finland, France, Germany, Jamaica, Ireland, Israel, United Kingdom, Luxembourg, Netherlands, Norway, Sweden, Switzerland, UAE, and Canada. **Center Recommended Depth - Consumers/User**
Lot, serial or model codes
Part number: 71026-70;  Lot number: 1281732; Exp. date: 2014/05; Lot number: 1283603; Exp. date: 2014/06;  Part number: 70827-71: Lot number: 1363015, Exp. Date: 2014/11; Lot number: 1363109, Exp. Date 2014/11:  Lot number: 1365921, Exp. Date 2014/12;  Part number 70827-72: Lot number: 1366006 Exp. Date : 2014/12; Lot number 1366111, Exp. Date: 2014/12; Lot number 1366337; Exp. Date: 2014/12; Lot number: 1366515, Exp Date: 2014/12;  Part number 70822-71; Lot number 1283345: Exp. Date: 2014/05; Lot number 1365920; Exp. Date: 2014/12; Lot number 1365934; Exp. Date: 2014/12;  Part number: 70822-72: Lot number: 1350414: Exp Date: 2014/07; Lot number 1363321; Exp. Date 2014/11:  Part number: 70819-70: Lot Number: 1365056: Exp. Date 2014/11; Lot number: 1366347: Exp. Date: 2014/12; Lot number: 1367917; Exp. Date 2014/12 Lot number: 1374907; Exp. Date 2015/03.
510(k) numbers
["K092602"]
PMA numbers
not on file
Product code
NBW
Firm FEI
3002803444
Firm city
Alameda
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register