FDA Medical Device Recalls (openFDA)

Covidien, LP: Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;

Recall number
Z-0485-2026
Recalling firm
Covidien, LP
Product
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
Status
Open, Classified
Initiated
2025-09-25
Posted
2025-11-07
Terminated
not on file
Reason
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
Root cause
Process control
Action
On September 25, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Customer actions: Identify and quarantine all unused impacted product listed. Return all quarantined devices to Medtronic. If purchased from a distributor, contact your distributor directly to arrange for the return of the devices back to your distributor. Please complete and return the enclosed Customer Confirmation Form. Pass on this notice to all those who need to be aware within your organization or to any organization where the listed affected lots may have been transferred or distributed.
Quantity
3263 units
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
Lot, serial or model codes
CFN: SIGSDL45CTVT; UDI-DIs: (1) 20884521741840, (2) 10884521741843; Lot numbers: N5F1971Y, N5F1723Y, N5D1858UY, N5B1134UY;
510(k) numbers
["K191070","K222641"]
PMA numbers
not on file
Product code
GDW
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register