FDA Medical Device Recalls (openFDA)
Abbott Diabetes Care, Inc.: FreeStyle Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count UPC 6 99073 12050 2; Freestyle Blood Glucose Test Strips; for Institutional Use only 50 count: UPC 6 99073 70792 5 Manufactured by Abbott Diabetes Care Inc. Alameda, CA. The FreeStyle Blood Glucose test strip is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger, upper arm and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.
- Recall number
- Z-0485-2014
- Recalling firm
- Abbott Diabetes Care, Inc.
- Product
- FreeStyle Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count UPC 6 99073 12050 2; Freestyle Blood Glucose Test Strips; for Institutional Use only 50 count: UPC 6 99073 70792 5 Manufactured by Abbott Diabetes Care Inc. Alameda, CA. The FreeStyle Blood Glucose test strip is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger, upper arm and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.
- Status
- Terminated
- Initiated
- 2013-11-18
- Posted
- 2013-12-17
- Terminated
- 2015-08-03
- Reason
- Certain lots of FreeStyle and FreeStyle Lite Blood Glucose Test Strips produce erroneously low blood glucose results when using FreeStyle Blood Glucose Meters, FreeStyle Flash Blood Glucose Meters and the FreeStyle blood glucose meter built into the OmniPod system. Erroneously low results that are not recognized may pose significant risks to your health.
- Root cause
- Other
- Action
- Abbott Diabetes Care sent an Urgent Product Recall letter in November 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers who were affected were instructed to take the following steps: Call Abbott Diabetes Care Customers Service immediately at 1-888-736-9869 for replacement of the affected product. While waiting for replacement strips to arrive, use an alternate method to measure their blood glucose. If only test strips available to customers were from affected lots, customers should not stop testing their blood glucose. Customers were instructed to dispose of the test strips in accordance with their local and state disposal regulations. Customers with questions were instructed to call Abbott Diabetes Care Customers Service at 1-888-736-9869. For questions regarding this recall call 510-239-2775. Abbott issued a press release at the request of CDRH on november 27, 2013.
- Quantity
- 141,219 cartons for human use - all lot numbers
- Distribution
- Worldwide Distribution - USA (nationwide and Puerto Rico) and Internationally to Australia, Austria, Belgium, Denmark, Egypt, Finland, France, Germany, Jamaica, Ireland, Israel, United Kingdom, Luxembourg, Netherlands, Norway, Sweden, Switzerland, UAE, and Canada. **Center Recommended Depth - Consumers/User**
- Lot, serial or model codes
- Part number 12050-21: Lot number: 1285007 Exp. date: 2014/06; Part number 70792-20 (instuitional use) Lot number: 1373262, Exp. Date 2015/02
- 510(k) numbers
- ["K092602"]
- PMA numbers
- not on file
- Product code
- NBW
- Firm FEI
- 3002803444
- Firm city
- Alameda
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28