Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Cardiovascular, LLC: Acrobat-i Positioner. Model Number C-XP-5000Z

Recall number
Z-0484-2025
Recalling firm
Maquet Cardiovascular, LLC
Product
Acrobat-i Positioner. Model Number C-XP-5000Z
Status
Open, Classified
Initiated
2024-10-16
Posted
2024-11-18
Terminated
not on file
Reason
Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
Root cause
Process change control
Action
MAQUET Cardiovascular LLC / Getinge notified consignees via letter FedEx 2-Day Delivery on 10/16/2024. Customers were instructed to forward the notification to all users and notify customers if further distributed, examine inventory and return any affected inventory, and to complete and return the MEDICAL DEVICE REMOVAL - RESPONSE FORM.
Quantity
14,338 units
Distribution
Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.
Lot, serial or model codes
Model Number C-XP-5000Z. UDI-DI Number: 00607567500006. Batch Numbers: 3000272452, 3000289270, 3000293675, 3000340401, 3000382289, 3000282355, 3000287205, 3000290368, 3000306921, 3000308238, 3000307458, 3000318694, 3000282741, 3000290350, 3000306394, 3000334128, 3000355318, 3000362345, 3000278266, 3000343467, 3000353846, 3000346421, 3000360292, 3000364706, 3000317546, 3000272461, 3000343116, 3000360175, 3000360569, 3000355540, 3000380141, 3000378810, 3000334322, 3000363049, 3000311731, 3000306920, 3000365139, 3000273855, 3000300092, 3000276339, 3000347351, 3000347358, 3000352249, 3000397151, 3000368288, 3000293958, 3000288356, 3000286557, 3000402809, 3000392590, 3000340972.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MWS
Firm FEI
2242352
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register