Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use Electrosurgical Snare Model/Catalog Number: SD-U400U-15 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.

Recall number
Z-0483-2026
Recalling firm
Olympus Corporation of the Americas
Product
Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use Electrosurgical Snare Model/Catalog Number: SD-U400U-15 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.
Status
Open, Classified
Initiated
2025-10-06
Posted
2025-11-07
Terminated
not on file
Reason
Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.
Root cause
Packaging
Action
On October 6, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: Examine your inventory and quarantine any affected devices. Cease usage of the product immediately. Please contact customer service to request a return material authorization, a credit will be issued for any returned product. Please forward this notice to other users who may have the affected products if you have further distributed it. Olympus requests you to report any complaints, including holes in device packaging, to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail Cynthia.Ow@Olympus.com.
Quantity
18174 units
Distribution
Worldwide distribution - US Nationwide and the countries of GU, and MP.
Lot, serial or model codes
Model: SD-400U-15; Lot Numbers: 27V, 28V, 29V, 2XV, 2YV, 2ZV, 31V, 32V, 33V, 34V, 35V, 36V, 37V, 38V, 39V, 3XV, 3YV, 3ZV, 41V, 42V, 43V, 44V, 45V, 46V, 47V, 48V, 49V, 4XV, 4YV, 4ZV, 51V, 52V, 53V, 54V, 55V, 56V UDI: 04953170408250
510(k) numbers
["K172734"]
PMA numbers
not on file
Product code
FDI
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register