Skip to the content

FDA Medical Device Recalls (openFDA)

Bayer Medical Care, Inc.: Brand Name: MEDRAD Stellant FLEX Syringe Kits Common Name: Angiographic Syringe Catalog/REF Number: FLEXD-150-SCS Description: 2 - 150 mL FLEX syringes, 1 large saline spike, 1 small contrast spike, patient t-connector tubing & prime tube. Catalog/REF Number: FLEXD-150-SPK Description: 2 - 150 mL FLEX syringes, 2 small spikes, patient t-connector tubing & prime tube.

Recall number
Z-0483-2021
Recalling firm
Bayer Medical Care, Inc.
Product
Brand Name: MEDRAD Stellant FLEX Syringe Kits Common Name: Angiographic Syringe Catalog/REF Number: FLEXD-150-SCS Description: 2 - 150 mL FLEX syringes, 1 large saline spike, 1 small contrast spike, patient t-connector tubing & prime tube. Catalog/REF Number: FLEXD-150-SPK Description: 2 - 150 mL FLEX syringes, 2 small spikes, patient t-connector tubing & prime tube.
Status
Terminated
Initiated
2020-10-05
Posted
not on file
Terminated
2021-03-19
Reason
Specific lots of MEDRAD Stellant FLEX 150 mL syringe kits may have holes and/or cracks that have the potential to allow fluid to leak out of the syringe.
Root cause
Process control
Action
On 10/08/020, Bayer sent all of their affected customers an Urgent Medical Device Field Corrective Action (recall) letter via e-mail. The recall letter requested that all of the affected customers immediately take the following steps: 1. Review your current inventory. Quarantine any affected batch/lot numbers referenced in the below chart. Instructions are included in this letter on how to identify batch/lot. 2. Complete the included response form and submit through email to CLSCRiCustomerLogistics@bayer.com or contact the Customer Logistics team at 724-940-7890. Please complete the Customer Response Form regardless of whether you have affected product or not. This action will assist us in tracking all affected product. 3. You will receive a Return Goods Authorization (RGA) number to return any affected Syringe kits remaining in your inventory for a credit. We can send new syringe kits with a separate purchase order as necessary. If product is needed immediately to avoid a stock out, expedited shipping options will be available. For expedited shipping options, please contact the Customer Logistics team at 724-940-7890. 4. Place the affected product in a box and label the outside of the container with the RGA number in large, bold writing. Use the shipping label provided via email when you receive the RGA number. 5. Upon Bayer receipt of return product, you will receive a credit on your account. Please distribute this letter to other appropriate departments and personnel within your institution who may have inventory that may be affected by this field corrective action.
Quantity
92,420 units
Distribution
Distributed to healthcare facilities nationwide throughout the US.
Lot, serial or model codes
Catalog/REF Number: FLEXD-150-SCS  Material Number: 87078817  Lot Numbers: 8584324, 8586383, 8586384, 8586385,   8586386, 8586388   UDI Numbers:   (01)00616258022888(11)200620(17)240620(10)8584324  (01)00616258022888(11)200620(17)240620(10)8584324  (01)00616258022888(11)200620(17)240620(10)8584324  (01)00616258022888(11)200620(17)240620(10)8584324  (01)00616258022888(11)200620(17)240620(10)8584324  (01)00616258022888(11)200620(17)240620(10)8584324  Catalog/REF Number: FLEXD-150-SPK Material Number: 85020218 Lot Number: 8585234  UDI Numbers: (01)00616258010656(11)200720(17)240720(10)8585234
510(k) numbers
["K192370"]
PMA numbers
not on file
Product code
DXT
Firm FEI
2520313
Firm city
Indianola
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register