Olympus Corporation of the Americas: Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.
- Recall number
- Z-0482-2026
- Recalling firm
- Olympus Corporation of the Americas
- Product
- Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.
- Status
- Open, Classified
- Initiated
- 2025-10-06
- Posted
- 2025-11-07
- Terminated
- not on file
- Reason
- Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.
- Root cause
- Packaging
- Action
- On October 6, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: Examine your inventory and quarantine any affected devices. Cease usage of the product immediately. Please contact customer service to request a return material authorization, a credit will be issued for any returned product. Please forward this notice to other users who may have the affected products if you have further distributed it. Olympus requests you to report any complaints, including holes in device packaging, to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail Cynthia.Ow@Olympus.com.
- Quantity
- 48208 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of GU, and MP.
- Lot, serial or model codes
- Model: SD-400U-10; Lot Numbers: 27V, 28V, 29V, 2XV, 2YV, 2ZV, 31V, 32V, 33V, 34V, 35V, 36V, 37V, 38V, 39V, 3XV, 3YV, 3ZV, 41V, 42V, 43V, 44V, 45V, 46V, 47V, 48V, 49V, 4XV, 4YV, 51V, 52V, 53V, 54V, 55V, 56V, 57V UDI: 04953170408243
- 510(k) numbers
- ["K172734"]
- PMA numbers
- not on file
- Product code
- FDI
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28