Accumetrics Inc: VerifyNow IIb/IIIa Test, 10-Test Kit, Catalog No. 85310, 10 tests/box; 690 total tests. The VerifyNow IIb/IlIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) lIb/IlIa receptor blockade in patients treated with abciximab or eptifibatide. VerifyNow lIb/IlIa Test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician. The VerifyNow System is a turbidimetric based optical detection system which measures platelet-induced aggregation. The system consists of an instrument, a disposable test device and quality control materials. The VerifyNow lIb/IlIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) lIb/IlIa receptor blockade in patients treated with abciximab or eptifibatide.

Recall number
Z-0482-2014
Recalling firm
Accumetrics Inc
Product
VerifyNow IIb/IIIa Test, 10-Test Kit, Catalog No. 85310, 10 tests/box; 690 total tests. The VerifyNow IIb/IlIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) lIb/IlIa receptor blockade in patients treated with abciximab or eptifibatide. VerifyNow lIb/IlIa Test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician. The VerifyNow System is a turbidimetric based optical detection system which measures platelet-induced aggregation. The system consists of an instrument, a disposable test device and quality control materials. The VerifyNow lIb/IlIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) lIb/IlIa receptor blockade in patients treated with abciximab or eptifibatide.
Status
Terminated
Initiated
2013-11-14
Posted
2013-12-11
Terminated
2013-12-23
Reason
Accumetrics is recalling the VerifyNow IIb/IIIa 10-Test because it contains incorrect information for the amount of time in which to perform the test following collection of the patient sample. The correct instruction is to perform the test within 15 minutes of collecting the patient sample.
Root cause
Other
Action
Accumetrics sent an Urgent Medical Device Correction letter dated November 14, 2013 to end users for the VerifyNow IIb/IIIa 10-Test. The letter informs the customers of the problems identified and the actions to be taken. The customers are instructed to call Accumetrics Customer Support at (800) 643-1640 option 2 or email at support@accumetrics.com. Customers are instructed to complete and return the enclosed Customer Account Tracking Form as soon as possible. For questions regarding this recall call 858-643-1600.
Quantity
69 units
Distribution
Nationwide Distribution including NY, PA, DC, CT, IN, NJ, AR, TX, AZ,and NM.
Lot, serial or model codes
Lot No. WC0181H, WC0181J, WC0181K
510(k) numbers
["K011337","K013596","K992531"]
PMA numbers
not on file
Product code
JOZ
Firm FEI
3001376288
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register