NeuroLogica Corporation: GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

Recall number
Z-0480-2026
Recalling firm
NeuroLogica Corporation
Product
GM85 Digital Mobile X-ray imaging System; Model Number: GM85;
Status
Open, Classified
Initiated
2025-10-07
Posted
2025-11-07
Terminated
not on file
Reason
The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.
Root cause
Device Design
Action
On October 7, 2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. The letters described the problem, product and actions to be taken. The customers were instructed to take the following Actions: Users may continue to use the GM85 system. Customers can inspect the unit for missing or loose bolts and if any issues with the bolts are detected, then please contact the service engineer immediately for inspection. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Complete and return the Customer Reply Form (CRF) Samsung Electronics will undertake the voluntary corrective action to reinforce the arm with a high tensile bolt to eliminate the risk of the THU falling and preventing any potential risk of injury. Service engineers will correct all affected devices free of charge and contact you to arrange for the correction. All corrections in the field started in September 2025 and will be completed by June 2026. If you have any questions, contact VP via email: ngujar@neurologica.com or call 978-564-8632.
Quantity
236 units
Distribution
Domestic: US Nationwide Distribution. International: Australia, Belgium, Canada, Chile, China Mainland, Czech Republic, Denmark, Ecuador, Finland, France, Germany, GHANA, Hungary, India, Italy, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab Emir., Vietnam,
Lot, serial or model codes
Model Number: GM85; UDI-DI: 08806088582412; Serial Numbers: 514GM3IK400001L, 5143M3IK300001X, 5143M3IJ100010M, 5143M3IHC00014H, 5143M3IK300008R, 5143M3JJ400007E, 5143M3HJ900002V, 5143M3IK400004F, 513YM3IK500003Z, 5143M3HK600005V, 5143M3JJ200010J, 5143M3HK600008M, 5143M3HK200004H, 5143M3JK500003W, 5143M3IJ200002F, 5143M3IHC00007A, 5143M3JJ300001K, 5143M3IJ600014H, 5143M3JK500006X, 5143M3JJ200017Y, 5143M3JK400014E, 5143M3IJ600013B, 5143M3GK500004A, 5143M3IHC00011V, 5143M3JK500001M, 5143M3JJ300012N, 5143M3GK500005M, 5143M3IK400007W, 5143M3JK400008N, 5143M3JJ300011Z, 5143M3JK100003Z, 5143M3JK500007L, 5143M3IK500001F, 5143M3IHC00002L, 5143M3IK300006T, 5143M3IHC00010R, 5143M3IK600004K, 5143M3IJ300003E, 5143M3IHC00003Y, 5143M3JJ400006K, 5143M3HJ900001L, 5143M3IJ600003L, 5143M3IJC00001Y, 5143M3IJ600001H, 5143M3HJ500003A, 5149M3IK400003E, 5143M3IJ800008L, 5143M3JJ700002R, 5143M3HK300002V, 5143M3JJ200008B, 513YM3IK500002R, 5143M3JK100005D, 5143M3IJ600004F, 5143M3IHC00005J, 5143M3JJC00004Y, 5143M3HK600003R, 5143M3JK500004D, 5143M3HK600004A, 5143M3HK300012L, 5143M3HK300008M, 5143M3JJ200015M, 5143M3HK300007N, 5143M3JK100007F, 5143M3IHC00001H, 5143M3IJ400006V, 5143M3IJ100002P, 514GM3HK100003J, 5143M3HK600010W, 5143M3HJ500001Y, 5143M3JK400010W, 5143M3IJ300005Z, 5143M3HK400010M, 5149M3IK400002K, 5143M3IK400002J, 5143M3GK500007W, 5143M3IK400009H, 5143M3IJ300002Y, 5143M3JK100004A, 513YM3IK400001W, 5143M3HK600011N, 5143M3HK300003Y, 5143M3JJ200012Z, 5143M3JK400007B, 5143M3HJA00002K, 5149M3IK400001H, 5143M3IJ100005R, 5143M3IHC00012Z, 5143M3HK400009W, 514GM3HK100002R, 5143M3IJ100001Y, 5143M3HK700002K, 5143M3IJ400004B, 5143M3JK400004M, 5143M3HJ900006H, 5143M3HK300005W, 5143M3JJ700001V, 5143M3IJ600015P, 5143M3JJ500002M, 5143M3HJ900003W, 5143M3IK600001Y, 5143M3JK100015B, 5143M3HJ600003K, 5143M3IJ200001D, 5143M3RJ100001M, 5143M3HK600009X, 5143M3IK600003N, 5143M3JJ300007T, 5143M3IJ400003R, 5143M3IK600002W, 5143M3HK200006K, 5143M3HJ900007R, 5143M3JK100002B, 5143M3IJ300006N, 5143M3IHC00017N, 5143M3IK300009L, 5143M3IJ500001B, 5143M3JJC00003L, 5143M3HJ500004M, 5143M3JK100006N, 5143M3HK700001D, 514GM3JK300001A, 5143M3IJ800005X, 5143M3HK400003P, 5143M3IK400005T, 5143M3HK300009K, 5143M3IJ600012X, 5143M3IHC00015L, 5143M3HK300010P, 5143M3RJ100003T, 5143M3DHB00001M, 5143M3IJ100003X, 5143M3IJ600006W, 5143M3IK400008A, 5143M3IHC00019T, 5143M3JK100014L, 513YM3IK400002A, 513YM3IK500005Y, 5143M3JK500002P, 5143M3HK400004J, 5143M3HJ500002J, 5143M3IJ400005N, 5143M3JJ400002X, 5143M3JJ200014F, 5143M3JJ700004K, 5143M3HK300017E, 5143M3IJ800011E, 5143M3HK300013A, 5143M3HK600007Z, 5143M3HK300004E, 5143M3GK500003J, 5143M3GK500001H, 5143M3GK500002Y, 5143M3GK500006P, 5143M3HK600002T, 5143M3HK600001D, 5143M3JK400013K, 5143M3HK400006F, 5143M3HK400007T, 5143M3HK400005Z, 5143M3HK400008X, 5143M3JK500005V, 513YM3IK500001K, 513YM3IK500004H, 5143M3HK300019Z, 5143M3JK400011A, 5143M3JK400003D, 5143M3JK400006R, 5143M3JK400009V, 5143M3JK400012H, 5143M3IK400003Z, 5143M3IK400006X, 5143M3HK300016Y, 5143M3IK300005H, 5143M3IK300007P, 5143M3HK300006Z, 5143M3HK300001F, 5143M3IK300002J, 5143M3HK200005V, 513YM3IK100002D, 513YM3IK100003N, 5143M3JK100013W, 514GM3HK100001E, 514GM3HK100004W, 5143M3JK100009M, 5143M3JK100008H, 5143M3JJC00006J, 5143M3JJC00002H, 5143M3HJA00010W, 5143M3HJA00008X, 513YM3IJA00001X, 5143M3HJA00007D, 5143M3HJA00001B, 5143M3HJ900010Z, 5143M3HJ900009E, 5143M3IJ800012D, 5143M3IJ800014B, 5143M3IJ800009M, 5143M3IJ800013Y, 514GM3JJ900001R, 5143M3IJ800007K, 5143M3IJ800001B, 5143M3IJ800002V, 5143M3IJ800003J, 5143M3IJ800004F, 5143M3JJ700003D, 5143M3JJ700005W, 5143M3HJ600005D, 5143M3IJ600009J, 5143M3IJ600008K, 5143M3IJ600007N, 5143M3IJ600005Y, 5143M3IJ600011M, 5143M3HJ500006W, 5143M3IJ600002P, 5143M3IJ400007P, 5143M3JJ400005H, 5143M3IJ300004W, 5143M3JJ300009R, 5143M3JJ300004B, 5143M3JJ300005D, 5143M3JJ200018X, 5143M3JJ300002X, 5143M3JJ300003J, 5143M3JJ200013D, 5143M3JJ200011L, 5143M3JJ200002M, 5143M3JJ200005X, 5143M3JJ200009A, 5143M3JJ200006R, 5143M3RJ100002V, 5143M3IJ100011V, 5143M3IJ100004E, 5143M3IHC00016Y, 5143M3IHC00009M, 5143M3IHC00008E, 5143M3IHC00006T;
510(k) numbers
["K222353"]
PMA numbers
not on file
Product code
IZL
Firm FEI
3004938766
Firm city
Danvers
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register