Enterix, Inc.: INSURE(R) QUIK FIT(TM) DEVELOPER KIT, Lot number B1478, Product Numbers: 30025 (InSure Quik FIT Developer Kit) and 50025 (INSURE(R) QUIK FIT(TM) containing Developer Kit and Patient Collection Kits. Product Number 30025 - InSure(R) Quik FIT)TM) Developer Kit (1 Box) . Each box contains: 25 InSure(R) Quik FIT(TM) Test Strips (1 per foil pack), 1 InSure(R) Quik FIT(TM) Conjugate Solution, I InSure(R) Quik FIT(TM) Run Buffer, and 1 InSure(R) Quik FIT(TM) Product Instructions. Product 50025 - InSure(R) Quik FIT(TM) contains 25 Patient Kits & 1 Developer Kit.

Recall number
Z-0480-2009
Recalling firm
Enterix, Inc.
Product
INSURE(R) QUIK FIT(TM) DEVELOPER KIT, Lot number B1478, Product Numbers: 30025 (InSure Quik FIT Developer Kit) and 50025 (INSURE(R) QUIK FIT(TM) containing Developer Kit and Patient Collection Kits. Product Number 30025 - InSure(R) Quik FIT)TM) Developer Kit (1 Box) . Each box contains: 25 InSure(R) Quik FIT(TM) Test Strips (1 per foil pack), 1 InSure(R) Quik FIT(TM) Conjugate Solution, I InSure(R) Quik FIT(TM) Run Buffer, and 1 InSure(R) Quik FIT(TM) Product Instructions. Product 50025 - InSure(R) Quik FIT(TM) contains 25 Patient Kits & 1 Developer Kit.
Status
Terminated
Initiated
2008-10-03
Posted
2008-12-29
Terminated
2009-03-26
Reason
False readings: Reduced sensitivity could affect the test line area and/or control line on the test strips, which may lead to inaccurate results. For further information, please contact Enterix, Inc. at 1-800-531--3681.
Root cause
Other
Action
Medical Device Recall Letters were sent on October 3, 2008 to all customers by certified mail. The letters were entitled "URGENT *** MEDICAL DEVICE RECALL (U.S. ONLY). Questions are to be directed to Enterix at 1-800-531-3681,
Quantity
45 kits (includes 25 patient collection kits and 1 developer kit)
Distribution
Product was distributed to medical facilities and laboratories throughout the US.
Lot, serial or model codes
Lot number B1478, Exp. Date 11/30/2008
510(k) numbers
["K060930"]
PMA numbers
not on file
Product code
KHE
Firm FEI
3003224146
Firm city
Edison
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register