Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, LP: EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.

Recall number
Z-0479-2021
Recalling firm
Encore Medical, LP
Product
EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.
Status
Open, Classified
Initiated
2020-06-19
Posted
not on file
Terminated
not on file
Reason
There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.
Root cause
Other
Action
The method of notification was via e-mail to surgical sales agents, and the notices were issued on 06/18/2020. Additionally, a notification to return the affected drills for replacement were sent via e-mail to affected surgical agents on 07/20/20. The following recommendations should be communicated to the implanting surgeons: Sales Agents Action: To avoid the potential binding, the following is recommended: " Use a plunge mill approach, i.e. plunge the drill in and out of the drill guide bushing a few times slowly until the hole is fully prepared. " Use of saline or fat to keep the drill lubricated during use. " Inspect all instrumentation prior to use to identify any wear, breakage or damage. Along with the notification, the communication includes an acknowledgment form that should be completed and returned to confirm that the instructions provided have been read and understood. THE ACKNOWLEDGEMENT SHOULD BE REVIEWED AND COMPLETED BY JULY 6, 2020.For questions, contact Teffany Hutto by phone at (512) 834-6255 or by email at teffany.hutto@djoglobal.com.
Quantity
28 units
Distribution
US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.
Lot, serial or model codes
all lot numbers
510(k) numbers
["K191325"]
PMA numbers
not on file
Product code
HSX
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06