Skip to the content

FDA Medical Device Recalls (openFDA)

Aesculap, Inc.: Aesculap GNI161 Bipolar Foot Control Product Usage: The Aesculap Bipolar Coagulator is intended for use in surgery to generate electrical power for bipolar instruments. Bipolar coagulators are used in neurosurgery, ENT surgery, urology, laparoscopy, and plastic surgery.

Recall number
Z-0479-2014
Recalling firm
Aesculap, Inc.
Product
Aesculap GNI161 Bipolar Foot Control Product Usage: The Aesculap Bipolar Coagulator is intended for use in surgery to generate electrical power for bipolar instruments. Bipolar coagulators are used in neurosurgery, ENT surgery, urology, laparoscopy, and plastic surgery.
Status
Terminated
Initiated
2013-09-10
Posted
2013-12-11
Terminated
2014-10-20
Reason
Complaints were received for the GN161 Bipolar Foot Control reporting that the bipolar energy did not stop after release of the foot control.
Root cause
Device Design
Action
Aesculap sent an Important Recall Notification letter dated September 23, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to return affected product and receive replacement product or credit. An Inventory Sheet is attached and must be completed. For questions call 610-984-9265 or 610-984-9414.
Quantity
85
Distribution
Worldwide Distribution - USA Nationwide and countries of: Austria, Germany, Japan, and Taiwan.
Lot, serial or model codes
all GN161's distributed between 4/28/2009 and 5/22/2013.
510(k) numbers
["K952524"]
PMA numbers
not on file
Product code
GEI
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register