Skip to the content

FDA Medical Device Recalls (openFDA)

Arjo, Inc.: BHM Kwiktrak Fixed Gate component of the Kwiktrak rail system used with the Maxi Sky 600 Ceiling Patient Lift; Manufactured by BMH Medical Inc., 2001 Tanguay Street , Magog, Quebec, Canada J1X 5Y5. The fixed gate, part #700.11550, was sold in two kits: a) Part #700.11500 Kwiktrak Gate Kit composed of the fixed gate and a 3-contact interface kit part #700.15515, and b) Part #700-11505 Double Gate Kit composed of two fixed gates and 3-contact interface kits

Recall number
Z-0479-2007
Recalling firm
Arjo, Inc.
Product
BHM Kwiktrak Fixed Gate component of the Kwiktrak rail system used with the Maxi Sky 600 Ceiling Patient Lift; Manufactured by BMH Medical Inc., 2001 Tanguay Street , Magog, Quebec, Canada J1X 5Y5. The fixed gate, part #700.11550, was sold in two kits: a) Part #700.11500 Kwiktrak Gate Kit composed of the fixed gate and a 3-contact interface kit part #700.15515, and b) Part #700-11505 Double Gate Kit composed of two fixed gates and 3-contact interface kits
Status
Terminated
Initiated
2006-10-23
Posted
2007-01-11
Terminated
2010-01-09
Reason
The Kwiktrak Gate stopper may fail and allow the ceiling lift to pass through to the end of the track and fall to the floor.
Root cause
Other
Action
Arjo Inc. USA sent recall letters to the direct accounts on 10/30/06, informing them that if the ceiling lift is moved too fast or too forcefully into a closed gate, the gate stopper can be damaged, causing its function to fail. One consequence can be that the gate may remain open, allowing the ceiling lift to pass through to the end of the track. The letters requested the customer to visually inspect their Kwiktrak Gate installation used with their Maxisky-600 Ceiling Lift to determine if there are any signs of impact or burr marks on the closed gate stopper. Illustrations of a damaged and normal gate stopper were included. The customers were requested to stop use of the Kwiktrak track system if any signs of impact or burr marks are found, and contact the Arjo Quality Department at 1-800-323-1245, ext. 6140 for special instructions. If no signs of damage were found, the customers were reminded to make sure that they manually drive the ceiling lift in a gentler manner to prevent damage to the gate stopper, making sure that all personnel using the ceiling lifts are properly trained. The accounts were informed that Arjo personnel will contact them in the next few weeks to install an improved system that will eliminate the risks of failure.
Quantity
3 gate systems
Distribution
Michigan, Ohio and Illinois
Lot, serial or model codes
Part #700.11550 installed in combination with part #700.15515.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSA
Firm FEI
1419652
Firm city
Roselle
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register