Stryker Sustainability Solutions: The Fuhrman Pleural/Pneumopericardial Drainage Set The Fuhrman Pleural/Pneumopericardial Drainage Set consists of pigtail catheter, access needle, dilator, and wire guide. The set also includes a multipurpose tubing adapter, and three-way stopcock.

Recall number
Z-0478-2016
Recalling firm
Stryker Sustainability Solutions
Product
The Fuhrman Pleural/Pneumopericardial Drainage Set The Fuhrman Pleural/Pneumopericardial Drainage Set consists of pigtail catheter, access needle, dilator, and wire guide. The set also includes a multipurpose tubing adapter, and three-way stopcock.
Status
Terminated
Initiated
2015-11-17
Posted
2016-01-04
Terminated
2016-09-28
Reason
Stryker Sustainability Solutions is recalling the Fuhrman Pleural & Pneumopericardial Drainage Set due to receiving complaints that it broke off in the pleural cavity while inserting the device into the patient.
Root cause
Device Design
Action
The firm, Stryker Sustainability Solutions, sent a "URGENT MEDICAL DEVICE RECALL" letter dated 11/17/2015 to its customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were instructed to discontinue the use of the product, and complete the Recall Effectiveness Check Form even if no product has been found in inventory. Customers should return the completed Recall Effectiveness Check Form to their local Stryker Sustainability Sales Representative, email to ssspfa@stryker.com, or mail to: Stryker Sustainability Solutions, 1810 West Drake Drive,Tempe, AZ 85283. If the firm indicates that affected devices remain in inventory, a prepaid shipping label will be issued for the return of the product. Customers will receive credit for all affected devices returned. Adverse reactions or quality problems experienced with the use of this product may be reported to: - Stryker Sustainability Solutions Complaint Hotline: +1(888) 888-3433x5555 - http://www.stryker.com/productexperience/ - The FDAs MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. If you have any questions, contact Senior Regulatory Affairs Specialist at (888) 888-3433x5506 or email: chelsea.cullen@stryker.com.
Quantity
34 devices
Distribution
US Distribution in the states of: AK, CA, WI, IN, FL, OR, NY, and NC.
Lot, serial or model codes
Item Number G03974  Lot Number 295397U 331621U 374044U 394431U 406125U 406131U 410982U 412655U 415094U 422060U 519812U 7718289U 838193U
510(k) numbers
not on file
PMA numbers
not on file
Product code
KDQ
Firm FEI
3002615969
Firm city
Phoenix
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register