Sonosite, Inc.: The software is ARM version 60.80.101.025 (referred to as Nano 1.0.5). The software is used with the SonoSite NanoMaxx ultrasound system.

Recall number
Z-0478-2011
Recalling firm
Sonosite, Inc.
Product
The software is ARM version 60.80.101.025 (referred to as Nano 1.0.5). The software is used with the SonoSite NanoMaxx ultrasound system.
Status
Terminated
Initiated
2010-08-13
Posted
2010-11-26
Terminated
2011-07-21
Reason
Software ARM version 60.80.101.025 (referred to as Nano1.0.5) on the NanoMaxx ultrasound system results in errors when the "print" command is selected. As a result, the displayed MI or TI values may be incorrect.
Root cause
Software design
Action
Sonosite issued Urgent Medical Device Correction letters dated August 9, 2010 to consignees. Consignees were advised of the software problem and given directions to verify the software version they are using. They are advised to contact SonoSite at 877-657-8118 to obtain a free software upgrade.
Quantity
69
Distribution
Worldwide Distribution: USA and to the countries of Australia, Great Britain, India, Israel, the Netherlands, New Zealand, Norway, and Spain.
Lot, serial or model codes
Part code: P11111-35. Serial numbers: 03HQF4, 03HQF3, 03HQF6, 03HQCM, 03HPZK, 03HQ9X, 03H9C3, 03HKN6,   03HPQ9, 03H87H, 03HQ9M, 03HQ9Q, 03HQB2, 03HQB3, 03HQBB, 03HQBX,   03HQBY, 03HQBZ, 03HQC2, 03HQC3, 03HQC4, 03HQC5, 03HQC6, 03HQCC,   03HQCD, 03HQCF, 03HQCG, 03HQCH, 03HQCJ, 03HQCK, 03HQCN, 03HQCP,   03HQCX, 03HQD2, 03HQD3, 03HQD4, 03HQD5, 03HQD8, 03HQD9, 03HQDB,   03HQDD, 03HQDF, 03HQDG, 03HQDH, 03HQDK, 03HQDL, 03HQDN,   03HQDP, 03HQFL, 03HQ77, 03H16Q, 03HKM7, 03HQBH, 03HQD6, 03HQCL,   03HQ6M, 03HQFG, 03HQF5, 03HQB9, 03HQDC, 03HQDT, 03HQDX, 03HQF0,   03HQF2, 03HQF7, 03HPL9, 03HQD7, 03HQC0, and 03HQFJ.
510(k) numbers
["K092058"]
PMA numbers
not on file
Product code
IYN
Firm FEI
3002087856
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register