Philips North America Llc: Wireless Footswtich used with the following Philips Allura Xper and Azurion Interventional Fluoroscopic X-ray System . Model Number: 722001, 722002, 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059, 722065, 722066, 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. Wireless footswitch Models: 459801319471, 459801319521, 459801319361, 459801319421, 459801319481, 459801319531, 459801319511, 459801319541, 459801319371, 459801319431, 459801319391, 459801319441, 459801238241, 459801238231, 459801238261, 459801238251, 459801238201, 459801238191, 459801238221, 459801238211, 459801733051, 459801733061, 459801733071, 459801733081, 459801257861, 459801733091

Recall number
Z-0476-2022
Recalling firm
Philips North America Llc
Product
Wireless Footswtich used with the following Philips Allura Xper and Azurion Interventional Fluoroscopic X-ray System . Model Number: 722001, 722002, 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059, 722065, 722066, 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. Wireless footswitch Models: 459801319471, 459801319521, 459801319361, 459801319421, 459801319481, 459801319531, 459801319511, 459801319541, 459801319371, 459801319431, 459801319391, 459801319441, 459801238241, 459801238231, 459801238261, 459801238251, 459801238201, 459801238191, 459801238221, 459801238211, 459801733051, 459801733061, 459801733071, 459801733081, 459801257861, 459801733091
Status
Open, Classified
Initiated
2021-10-29
Posted
2022-01-10
Terminated
not on file
Reason
Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure
Root cause
Device Design
Action
Philips notified customers via an Urgent Medical Device Correction letter on October 29th, 2021 via Certified Mail. Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk and actions that should be taken by the user in order to prevent risks for patients or users " Ensure that the wired foot switch is always connected to the system. The wired foot switch is always provided with the Allura Xper and the Azurion systems. " Immediately start using the wired foot switch in case the wireless foot switch loses connection. " Circulate this notice to all users of this device so that they are aware of the issue. " Place this Medical Device Correction with the documentation of the Philips Allura Xper or Azurion system. " Return the attached reply form to Philips to confirm that: the Allura Xper and Azuarion system wired foot switch is connected to the system the users of the system have reviewed and understood this Medical Device Correction. S. The actions planned by Philips IGT Systems to correct the problem You will be contacted by your local Philips representative as soon as this solution is available. If you need any further information or support concerningthis issue, please contact your local Philips representative. Technical Support Line: 1-800-722-9377
Quantity
1971 Total; USA: 681 systems R.O.W.: 1,290 systems
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Bangladesh, Belgium, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Korea, Republic of Kuwait, Latvia, Lebanon, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, State of Philippines, Poland, Portugal, R¿union, Romania, RUMANIA, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, UZBEKISTAN, Viet Nam.
Lot, serial or model codes
Wireless foot switches Models: WFS Ad7(X) Option 3P 459801319471; WFS Ad7(X) Option 4P+2 459801319521; WFS Ad7(X) Option 3P 459801319361; WFS Ad7(X) Option 4P+2 459801319421; WFS Field Extension Set 3P (ORT) 459801319481; WFS Field Extension Set 4P+2 (ORT) 459801319531; WFS Field Extension Set 3P 459801319511; WFS Field Extension Set 4P+2 459801319541; WFS Field Extension Set Mono 459801319371; WFS Field Extension Set Biplane 459801319431; WFS Field Extension Set Mono (ORT) 459801319391; WFS Field Extension Set Biplane (ORT) 459801319441; Wireless Footswitch Set 3P 459801238241; Wireless Footswitch 3P 459801238231; Wireless Footswitch Set 4P+2 459801238261; Wireless Footswitch 4P+2 459801238251 Wireless Footswitch Set JP 459801238201; Wireless Footswitch 3P 459801238191; Wireless Footswitch Set 4P+2 459801238221; Wireless Footswitch 4P+2 459801238211; Wireless Footswitch 3P Kit 459801733051; Wireless Footswitch 3P Kit 459801733061; Wireless Footswitch 4P+2 Kit 459801733071; Wireless Footswitch 4P+2 Kit 459801733081; WFS Base Station 459801257861; WFS Base Station Kil 459801733091.
510(k) numbers
K161563, K200917
PMA numbers
not on file
Product code
OWB
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source