FDA Medical Device Recalls (openFDA)
Arjo, Inc. dba ArjoHuntleigh: Enterprise 8000 Bed; an AC-powered adjustable hospital bed; ArjoHuntleigh AB, Eslov, Sweden; Models 8000, 8001, 8002, 8003, 8500, 8600, 8700 and 8800 General medicine, general and specialist surgery, care of the elderly, HDU, ITU, coronary care and oncology. The bed is appropriate for high dependency patients who pose a movement and handling risk and/or whose clinical condition requires that they are positioned with the minimal physical handling.
- Recall number
- Z-0476-2013
- Recalling firm
- Arjo, Inc. dba ArjoHuntleigh
- Product
- Enterprise 8000 Bed; an AC-powered adjustable hospital bed; ArjoHuntleigh AB, Eslov, Sweden; Models 8000, 8001, 8002, 8003, 8500, 8600, 8700 and 8800 General medicine, general and specialist surgery, care of the elderly, HDU, ITU, coronary care and oncology. The bed is appropriate for high dependency patients who pose a movement and handling risk and/or whose clinical condition requires that they are positioned with the minimal physical handling.
- Status
- Terminated
- Initiated
- 2012-11-16
- Posted
- 2013-08-06
- Terminated
- 2016-09-13
- Reason
- ArjoHuntleigh has received reports of unintended movement where the backrest and/or thigh sections of the bed have moved without anyone consciously pressing the control button. These incidents have occurred in High Dependency and Intensive Therapy environments where unintended movement of the patient may have serious consequences.
- Root cause
- Under Investigation by firm
- Action
- Please be aware that this is not a new recall. Posting to the Internet was delayed due to a technical error. ArjoHuntleigh sent an Urgent Field Correction Recall letters dated November 16, 2012, to all affected customers, informing the accounts of reports of unintended movement of the backrest and/or thigh sections of the bed without intentional activation of the control buttons, which could have serious consequences due to the patient's links to other life supporting devices. The accounts were requested to disseminate the information to all users of the affected beds, to attach the enclosed warning labels to the outside of both head end safety sides immediately adjacent to the controls, to include a copy of the notice in the Instruction for Use Manuals for their beds, and to complete the Customer Response Form online, using the enclosed card's instructions. Questions were directed to the ArjoHuntleigh at 1-800-323-1245, ext 57985.
- Quantity
- 94 beds in U.S., 31 beds in Canada, 26,010 beds in ROW
- Distribution
- Worldwide Distribution-USA (nationwide, including Puerto Rico) and the countries of Algeria, Angola, Australia, Austria, Bahrain, Belgium, Bosnia and Herzegovia, Botswana, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Ecuador, Egypt, Falkland Islands, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Libyan Arab Jamahira, Lithuania, Malaysia, Malta, Mauritius, Mexico, Morocco, the Netherlands, New Zealand, Nigeria, Nordic, Norway, Oman, Panama, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Switzerland, Syrian Arab Republic, Thailand, Turkey, United Kingdom, United Arab Emirates, Uruguay, Venezuela and Yemen.
- Lot, serial or model codes
- Models 8000, 8001, 8002, 8003, 8500, 8600, 8700 and 8800; all serial numbers
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FNL
- Firm FEI
- 1419652
- Firm city
- Addison
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28