FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: Baxter Interlink System 3-Port Manifold with Check Valves, Vol. 1.0 mL, Large Bore 4-Way Stopcock Extension Set, Vol. 6.2 mL, 41" (104 cm), 2 Injection Sites, Male Luer Lock Adapter; an Rx sterile, nonpyrogenic fluid pathway for the administration of sterile I.V. fluids to the patient; Baxter Healthcare Corporation, Deerfield, IL 60015 USA; product code 2C6921 Intended use: for the administration of sterile I.V. fluids to the patient
- Recall number
- Z-0476-2012
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Baxter Interlink System 3-Port Manifold with Check Valves, Vol. 1.0 mL, Large Bore 4-Way Stopcock Extension Set, Vol. 6.2 mL, 41" (104 cm), 2 Injection Sites, Male Luer Lock Adapter; an Rx sterile, nonpyrogenic fluid pathway for the administration of sterile I.V. fluids to the patient; Baxter Healthcare Corporation, Deerfield, IL 60015 USA; product code 2C6921 Intended use: for the administration of sterile I.V. fluids to the patient
- Status
- Terminated
- Initiated
- 2011-12-06
- Posted
- 2012-01-11
- Terminated
- 2012-11-01
- Reason
- During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection.
- Root cause
- Employee error
- Action
- The firm, Baxter, sent an "Urgent Product Recall" letter dated December 6, 2011, via first class mail to its customers. The letter described the product, problem, and actions to be taken. The customers were instructed to inspect all locations where their facility's inventory is located for products with the affected lot numbers; return the affected product; request a credit or replacement by calling Baxter Healthcare Center for Service at 1-888-229-0001 between 7:00am to 6:00pm Central time, Monday through Friday; complete and return the attached Customer Reply Form via fax to: 1-847-270-5457, or scan and email to fca@baxter.com; and if they are a dealer, wholesaler, or distributor/reseller that distributed any affected product to other facilities, please notify their customers of this recall. If you have any questions regarding the communication, please call The Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.
- Quantity
- 3,216 units
- Distribution
- Worldwide distribution: USA (nationwide) including states of: Alaska, California, Colorado, Florida, Georgia, Hawaii, Illinois, Kansas, Kentucky, Massachusetts, Maine, Michigan, Minnesota, Missouri, North Carolina, New Hampshire, New Jersey, New York, Ohio, Pennsylvania, Rhode Island, South Carolina, Tennessee, Virginia, Washington, Wisconsin and West Virginia; and country of: New Zealand.
- Lot, serial or model codes
- product code 2C6921, lot number GR295634
- 510(k) numbers
- ["K932512"]
- PMA numbers
- not on file
- Product code
- FMG
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28