FDA Medical Device Recalls (openFDA)
Heartware, Inc.: HVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT, Model Number 1100
- Recall number
- Z-0475-2024
- Recalling firm
- Heartware, Inc.
- Product
- HVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT, Model Number 1100
- Status
- Open, Classified
- Initiated
- 2023-10-31
- Posted
- 2023-12-02
- Terminated
- not on file
- Reason
- Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended troubleshooting actions and (2) instructions relating to the useful life of the HVAD system components
- Root cause
- Labeling design
- Action
- Heartware issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 10/31/2023 via letter (UPS). The notice explained the issue and informed them of the updates to the HVAD IFU and PM (patient manual).
- Quantity
- 45 units
- Distribution
- Worldwide
- Lot, serial or model codes
- Serial numbers: HW001, HW012, HW016, HW017, HW023, HW025, HW028, HW044, HW045, HW046, HW047, HW049, HW050, HW051, HW061, HW064, HW083, HW084, HW086, HW091, HW094, HW104, HW105, HW108, HW109, HW110, HW114, HW115, HW116, HW119, HW126, HW127, HW129, HW130, HW131, HW138, HW139, HW143, HW146, HW147, HW153, HW156, HW161, HW170, HW181
- 510(k) numbers
- not on file
- PMA numbers
- ["P100047"]
- Product code
- DSQ
- Firm FEI
- 3007042319
- Firm city
- Miami Lakes
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28