Skip to the content

FDA Medical Device Recalls (openFDA)

Arthrosurface, Inc.: Total Toe Instrument Kit, Labeled in part: "Instrument Set, 15mm Toe, Dorsal Flange Combi" Part Number: 9000-3002 The component in question is the hex driver (part number 2009-4001).

Recall number
Z-0475-2016
Recalling firm
Arthrosurface, Inc.
Product
Total Toe Instrument Kit, Labeled in part: "Instrument Set, 15mm Toe, Dorsal Flange Combi" Part Number: 9000-3002 The component in question is the hex driver (part number 2009-4001).
Status
Terminated
Initiated
2014-06-09
Posted
2015-12-23
Terminated
2016-02-05
Reason
An oversight in the implementation of the design change record resulted in outdated components being shipped to customers. The component did not pass a half-cycle sterilization.
Root cause
Process change control
Action
The distributors were verbally notified, and replacement Rev F Hex Drivers sent to the locations of the kits. The kits had been moved in preparation for upcoming surgeries, but the Drivers were replaced prior to any additional surgery taking place. No affected product remains in the field. For further questions, please call (508) 520-3003.
Quantity
3 units
Distribution
US Distribution to the states of : PA, WV and CA.
Lot, serial or model codes
SN 000242 SN 000244 SN 000246 SN 000247
510(k) numbers
["K132496"]
PMA numbers
not on file
Product code
LZJ
Firm FEI
3004154314
Firm city
Franklin
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register