Skip to the content

FDA Medical Device Recalls (openFDA)

ConMed Corporation: Foot Controlled Electrosurgical Suction Coagulator, Electrosurgical Suction Coagulator 10FR, Catalog Number 130187

Recall number
Z-0474-2023
Recalling firm
ConMed Corporation
Product
Foot Controlled Electrosurgical Suction Coagulator, Electrosurgical Suction Coagulator 10FR, Catalog Number 130187
Status
Terminated
Initiated
2022-11-18
Posted
2022-12-09
Terminated
2026-04-13
Reason
Suction ports may potentially be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use, which could cause a delay in procedure and could require another device to be used.
Root cause
Process control
Action
On November 18, 2022, the firm notified affected customers through Urgent Medical Device Recall letters. Customers were instructed to review their inventory for any of the affected product and return affected product to the CONMED. Open or used devices should not be returned. If you have any questions or requests, please contact the Field Action Support Team at 1-800-448-6506 (8:00am to 7:00pm EST Monday through Friday) or email SC130187@conmed.com.
Quantity
23,700 US; 5,225 OUS
Distribution
Domestic distribution nationwide. Foreign distribution to Colombia, Ecuador, Nicaragua, Panama, Thailand, Australia, Canada, and Lebanon.
Lot, serial or model codes
UDI-DI: 30653405059930 Lot codes 202204294 202205134 202205164 202206084 202206154 202206204 202206304 202207074 202207084 202208014 202208044 202208084
510(k) numbers
["K033003"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3001451571
Firm city
Utica
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register