Becton Dickinson & Co.: Vancomycin Screen Agar
- Recall number
- Z-0473-2016
- Recalling firm
- Becton Dickinson & Co.
- Product
- Vancomycin Screen Agar
- Status
- Terminated
- Initiated
- 2015-06-29
- Posted
- 2015-12-23
- Terminated
- 2016-08-12
- Reason
- The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life.
- Root cause
- Under Investigation by firm
- Action
- Becton Dickinson (BD) sent US customers a Urgent Product Recall letter dated June 2015, via UPS. Customers were requested to return response forms indicating acknowledgement of the recall and quantity of remaining product that was at their facility. The acknowledgement form included with the follow up letter that is faxed back to BD will be reconciled with the customer list. BD will contact 10% of non-respondents to verify receipt of the customer letter. Any product returned to BD will be quarantined and discarded. If further assistance is needed, please contact BD Technical Services at 1-800-638-8663.
- Quantity
- 60,840
- Distribution
- Worldwide Distribution - US (nationwide) Distribution to the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI and WV., and to the countries of : Canada, Colombia, India, Mexico and Europe.
- Lot, serial or model codes
- catalog number 222204. Lot numbers 5041659, 5064921, 5076659, 5083712, 5097504, 5105955, 5130907, 5139626, and 5147941
- 510(k) numbers
- ["K951539"]
- PMA numbers
- not on file
- Product code
- JWY
- Firm FEI
- 1119779
- Firm city
- Sparks
- Firm state
- MD
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28